12 months from the date of the invoice, subject to the following conditions: In case of justified complaints, KaVo will honour its warranty with a free replace‐ ment or repair. Other claims of any nature whatsoever, in particular with respect to compensation, are excluded.
Page 5
Instructions for use GENTLEsilence 8000BS - 1.000.5600 1 User instructions | 1.1 Warranty terms and conditions All liability is excluded if defects or their consequences originate from manipula‐ tions or changes to the product made by the customer or a third party that is not authorised by KaVo.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions Warning symbol Structure DANGER The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 2 Safety | 2.3 Safety instructions According to these regulations, this medical device may only be used for the de‐ scribed application by a knowledgeable user. The following must be observed: ▪ the applicable health and safety regulations ▪...
Page 8
This service interval depends on the frequency of use and should be adjusted accordingly. Service may only be carried out by KaVo-trained repair shops using original KaVo replacement parts. 6/19...
Instructions for use GENTLEsilence 8000BS - 1.000.5600 3 Product description | 3.1 Technical data 3 Product description Mat. no. 1.000.5600 GENTLEsilence LUX 8000 BS ( 3.1 Technical data Minimum pressure 2.7 bar Drive pressure 2.7 to 3.2 bar recommendation 2,8 bar Air consumption 47 to 53 Nl/min.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 4 First use | 4.1 Check the amount of water 4 First use WARNING Hazard from non-sterile products. Infection hazard for care provider and patient. ▶ Before first use and after each use, prepare and sterilise the medical device and accessories accordingly.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 5 Operation | 5.1 Attach the medical device 5 Operation Note At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 min. without the instrument being attached; if there is a risk of contamination from reflux or back suction, the system must be rinsed for 20-30 seconds.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 5 Operation | 5.4 Removing the milling tool or diamond grinder CAUTION Injury from improper use of carbide cutters or diamond grinders. Never use metal cutters or grinders with worn shafts can fall out during treatment and injure the patient.
Page 13
Instructions for use GENTLEsilence 8000BS - 1.000.5600 5 Operation | 5.4 Removing the milling tool or diamond grinder WARNING Do not press the pushbutton while the cutter or grinder is rotating. If you press the pushbutton when the cutter or grinder is rotating, it can damage the chucking system and cause injury.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 6 Troubleshooting | 6.1 Exchanging the O-rings on the coupling on the supply hose 6 Troubleshooting 6.1 Exchanging the O-rings on the coupling on the supply hose CAUTION Hazard from improper care of the O-rings.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 7 Preparation methods according to ISO 17664 | 7.1 Preparation at the site of use 7 Preparation methods according to ISO 17664 WARNING Hazard from non-sterile products. Infection hazard for care provider and patient.
® ® only refers to the material compatibility with KaVo products.) ▶ For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector. ▶ Directly after automated cleaning, treat the medical device with the care prod‐...
Reduced product life. ▶ Perform proper care regularly! Note KaVo only guarantees that its products will function properly when the care prod‐ ucts used are those listed as accessories, as they were tested for proper use on our products. 15/19...
▶ Remove the cutter or grinder, place the spray nipple tip in the opening and spray. Note Carry out servicing according to instructions in the section "Care with KaVo Spray". 7.5.2 Care with KaVo QUATTROcare Cleaning and care unit with expansion pressure for effective cleaning and care.
Sterilisation in a steam steriliser (Autoclave) EN 13060/ISO 17665-1 CAUTION Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Before each sterilisation cycle, service the medical device with KaVo care products. CAUTION Contact corrosion due to moisture.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 7 Preparation methods according to ISO 17664 | 7.8 Storage 7.8 Storage Prepared products must be stored, protected from germs (as far as possible) and dust, in a dry, dark, cool room. Note Comply with the expiry date of the sterilised items.
Instructions for use GENTLEsilence 8000BS - 1.000.5600 8 Accessories 8 Accessories Accessories obtainable from dental and medical suppliers. Material summary Mat. no. Replacement turbine with key 1.003.6883 Replacement turbine without key 1.003.6884 Cleanpac 10 units 0.411.9691 STERIclave bags 0.411.9912 Nozzle needle 0.410.0921...