Summary of Contents for KaVo GENTLEsilence 8000BN

  • Page 1 Instructions for use GENTLEsilence 8000BN - 1.000.5700 Always be on the safe side.
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 (0) 7351 56-0 www.kavo.com Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 Contents Contents User instructions ............................2 Warranty terms and conditions ......................2 Safety ................................ 4 Description of safety instructions ..................... 4 Purpose – Proper use ........................4 Safety instructions ........................... 5 Product description ............................ 7 Technical data ..........................
  • Page 4: User Instructions

    12 months from the date of the invoice, subject to the following conditions: In case of justified complaints, KaVo will honour its warranty with a free replace‐ ment or repair. Other claims of any nature whatsoever, in particular with respect to compensation, are excluded.
  • Page 5 Instructions for use GENTLEsilence 8000BN - 1.000.5700 1 User instructions | 1.1 Warranty terms and conditions All liability is excluded if defects or their consequences originate from manipula‐ tions or changes to the product made by the customer or a third party that is not authorised by KaVo.
  • Page 6: Safety

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions Warning symbol Structure DANGER The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance.
  • Page 7: Safety Instructions

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 2 Safety | 2.3 Safety instructions According to these regulations, this medical device may only be used for the de‐ scribed application by a knowledgeable user. The following must be observed: ▪ the applicable health and safety regulations ▪...
  • Page 8 This service interval depends on the frequency of use and should be adjusted accordingly. Service may only be carried out by KaVo-trained repair shops using original KaVo replacement parts. 6/19...
  • Page 9: Product Description

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 3 Product description | 3.1 Technical data 3 Product description Mat. no. 1.000.5700 GENTLEsilence LUX 8000 BN ( 3.1 Technical data Minimum pressure 2.7 bar Drive pressure 2.7 to 3.2 bar recommendation 2.8 bar Air consumption 47 to 53 Nl/min.
  • Page 10: First Use

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 4 First use | 4.1 Check the amount of water 4 First use WARNING Hazard from non-sterile products. Infection hazard for care provider and patient. ▶ Before first use and after each use, prepare and sterilise the medical device and accessories accordingly.
  • Page 11: Operation

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 5 Operation | 5.1 Attach the medical device 5 Operation Note At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 min. without the instrument being attached; if there is a risk of contamination from reflux or back suction, the system must be rinsed for 20-30 seconds.
  • Page 12: Removing The Milling Tool Or Diamond Grinder

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 5 Operation | 5.4 Removing the milling tool or diamond grinder CAUTION Injury from using worn cutters or grinders. Cutters or grinders can fall out during treatment and injure the patient. ▶ Never use cutters or grinders with worn shafts.
  • Page 13 Instructions for use GENTLEsilence 8000BN - 1.000.5700 5 Operation | 5.4 Removing the milling tool or diamond grinder WARNING Do not press the pushbutton while the cutter or grinder is rotating. If you press the pushbutton when the cutter or grinder is rotating, it can damage the chucking system and cause injury.
  • Page 14: Troubleshooting

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 6 Troubleshooting | 6.1 Exchange the O-rings on the medical device 6 Troubleshooting 6.1 Exchange the O-rings on the medical device Note The O-ring on the coupling may only be lubricated with a cotton ball wet with KA‐...
  • Page 15 Instructions for use GENTLEsilence 8000BN - 1.000.5700 6 Troubleshooting | 6.3 Change the water filter ▶ Pull back the insert ② slightly. ▶ Place the sleeve ① on the insert ②, and screw it tight clockwise. 13/19...
  • Page 16: Preparation Methods According To Iso 17664

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 7 Preparation methods according to ISO 17664 | 7.1 Preparation at the site of use 7 Preparation methods according to ISO 17664 WARNING Hazard from non-sterile products. Infection hazard for care provider and patient.
  • Page 17: Mechanically Cleaning The Exterior And Interior

    ® ® only refers to the material compatibility with KaVo products.) ▶ For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector. ▶ Directly after automated cleaning, treat the medical device with the care prod‐...
  • Page 18: Mechanically Disinfecting The Exterior And Interior

    Reduced product life. ▶ Perform proper care regularly! Note KaVo only guarantees that its products will function properly when the care prod‐ ucts used are those listed as accessories, as they were tested for proper use on our products. 16/19...
  • Page 19: Care With Kavospray

    7 Preparation methods according to ISO 17664 | 7.6 Packaging 7.5.1 Care with KAVOspray KaVo recommends servicing the project twice daily (at noon and in the evening after hours), after each time the machine is cleaned, and before each sterilisation.
  • Page 20: Sterilisation

    Sterilisation in a steam steriliser (Autoclave) EN 13060/ISO 17665-1 CAUTION Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Before each sterilisation cycle, service the medical device with KaVo care products. CAUTION Contact corrosion due to moisture.
  • Page 21: Accessories

    Instructions for use GENTLEsilence 8000BN - 1.000.5700 8 Accessories 8 Accessories Accessories obtainable from dental and medical suppliers. Material summary Mat. no. Replacement turbine with key 1.003.6883 Replacement turbine without key 1.003.6884 Cleanpac 10 units 0.411.9691 STERIclave bags 0.411.9912 O-ring 7.4 x 0.8 1.005.8152...