Instructions for use GENTLEsilence 8000BN - 1.000.5700 1 User instructions | 1.1 Warranty terms and conditions All liability is excluded if defects or their consequences originate from manipula‐ tions or changes to the product made by the customer or a third party that is not authorised by KaVo.
Instructions for use GENTLEsilence 8000BN - 1.000.5700 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions Warning symbol Structure DANGER The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance.
Instructions for use GENTLEsilence 8000BN - 1.000.5700 2 Safety | 2.3 Safety instructions According to these regulations, this medical device may only be used for the de‐ scribed application by a knowledgeable user. The following must be observed: ▪ the applicable health and safety regulations ▪...
Page 8
This service interval depends on the frequency of use and should be adjusted accordingly. Service may only be carried out by KaVo-trained repair shops using original KaVo replacement parts. 6/19...
Instructions for use GENTLEsilence 8000BN - 1.000.5700 3 Product description | 3.1 Technical data 3 Product description Mat. no. 1.000.5700 GENTLEsilence LUX 8000 BN ( 3.1 Technical data Minimum pressure 2.7 bar Drive pressure 2.7 to 3.2 bar recommendation 2.8 bar Air consumption 47 to 53 Nl/min.
Instructions for use GENTLEsilence 8000BN - 1.000.5700 4 First use | 4.1 Check the amount of water 4 First use WARNING Hazard from non-sterile products. Infection hazard for care provider and patient. ▶ Before first use and after each use, prepare and sterilise the medical device and accessories accordingly.
Instructions for use GENTLEsilence 8000BN - 1.000.5700 5 Operation | 5.1 Attach the medical device 5 Operation Note At the beginning of each workday, the water-conducting systems should be rinsed for at least 2 min. without the instrument being attached; if there is a risk of contamination from reflux or back suction, the system must be rinsed for 20-30 seconds.
Instructions for use GENTLEsilence 8000BN - 1.000.5700 5 Operation | 5.4 Removing the milling tool or diamond grinder CAUTION Injury from using worn cutters or grinders. Cutters or grinders can fall out during treatment and injure the patient. ▶ Never use cutters or grinders with worn shafts.
Page 13
Instructions for use GENTLEsilence 8000BN - 1.000.5700 5 Operation | 5.4 Removing the milling tool or diamond grinder WARNING Do not press the pushbutton while the cutter or grinder is rotating. If you press the pushbutton when the cutter or grinder is rotating, it can damage the chucking system and cause injury.
Instructions for use GENTLEsilence 8000BN - 1.000.5700 6 Troubleshooting | 6.1 Exchange the O-rings on the medical device 6 Troubleshooting 6.1 Exchange the O-rings on the medical device Note The O-ring on the coupling may only be lubricated with a cotton ball wet with KA‐...
Page 15
Instructions for use GENTLEsilence 8000BN - 1.000.5700 6 Troubleshooting | 6.3 Change the water filter ▶ Pull back the insert ② slightly. ▶ Place the sleeve ① on the insert ②, and screw it tight clockwise. 13/19...
Instructions for use GENTLEsilence 8000BN - 1.000.5700 7 Preparation methods according to ISO 17664 | 7.1 Preparation at the site of use 7 Preparation methods according to ISO 17664 WARNING Hazard from non-sterile products. Infection hazard for care provider and patient.
® ® only refers to the material compatibility with KaVo products.) ▶ For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector. ▶ Directly after automated cleaning, treat the medical device with the care prod‐...
Reduced product life. ▶ Perform proper care regularly! Note KaVo only guarantees that its products will function properly when the care prod‐ ucts used are those listed as accessories, as they were tested for proper use on our products. 16/19...
7 Preparation methods according to ISO 17664 | 7.6 Packaging 7.5.1 Care with KAVOspray KaVo recommends servicing the project twice daily (at noon and in the evening after hours), after each time the machine is cleaned, and before each sterilisation.
Sterilisation in a steam steriliser (Autoclave) EN 13060/ISO 17665-1 CAUTION Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Before each sterilisation cycle, service the medical device with KaVo care products. CAUTION Contact corrosion due to moisture.
Instructions for use GENTLEsilence 8000BN - 1.000.5700 8 Accessories 8 Accessories Accessories obtainable from dental and medical suppliers. Material summary Mat. no. Replacement turbine with key 1.003.6883 Replacement turbine without key 1.003.6884 Cleanpac 10 units 0.411.9691 STERIclave bags 0.411.9912 O-ring 7.4 x 0.8 1.005.8152...