This product is intended for use by industrial customers and should be serviced only by Brooks or Brooks trained representatives. The service manuals and related materials are provided in English at no charge and are intended for use by experienced technicians.
The IntelliXcap can only be used with tubes and cartridges that have been configured and tested. Do not use alternative tubes and cartridges that have not been configured and tested. Untrained or Improperly Equipped Personnel Untrained or improperly equipped personnel performing this procedure may cause damage to the equipment.
Page 13
Chemical Hazard The IntelliXcap may be used to de-/recap samples that expose users to chemical hazards which, if not properly handled, may result in death or serious injury. Read and understand the safety information for the equipment where the IntelliXcap is used.
There is a safety switch installed on the access door that, if the door is not completely closed, the IntelliXcap will not function. Opening of the access door while tool is in use initiates an immediate stop function of all moving parts.
Page 15
41 for further information on using the IntelliXcap within an Integrated system. When the IntelliXcap is commissioned and starts functioning, the automated door will close, the status message will appear on the operator monitor: Initializing please wait The safety door must be activated and tested before commissioning the machine for use.
The IntelliXcap can be controlled manually or can be integrated into an automated robotic system using a serial command set. The IntelliXcap can work with a range of tube types – a specific cartridge is required for each different cap design format. The current list of available...
This manual contains information on safety, specifications, and operation as well as troubleshooting and maintenance of the IntelliXcap. If there are any questions regarding this manual or use of this system or to order additional copies of this publication, contact Brooks Automation Service. See "For Technical Support:"...
Ensure enough space is available to accommodate the ejected stage platform. The machine has a square footprint and is regarded as highly stable. Place the IntelliXcap in a well-ventilated area on top of an even surface, solid enough to carry its weight. The surface must comply with 1.3.1 of Annex I of 2006/42/EC.
Part Number: 319430 Rev. A Site Requirements Environmental Requirements The IntelliXcap shall be used within the rule set of the Good Laboratory Practices, GLP. The machine must be operated indoors and under the following environmental specifications only: Table 3-3: Environmental Requirements...
Move the kit to the appropriate unpack area. Unpack the kit and inspect and confirm the contents. Report any missing or damaged items to Brooks Automation. Review this procedure and confirm that you have the proper items required to do the job.
Page 27
Remove the shipping tape securing both the door (seen above) and the stage seen to the left. Keep the original packing material in a dry/low humidity location in case the IntelliXcap needs to be transported for service or repair. Follow all local regulations while disposing the original packing solution.
IntelliXcap is covered for both normal operating conditions and emergency conditions. The IntelliXcap shall be screened off appropriately before operation and it must only be used to remove and replace screw caps on micro tubes in specific, SBS-footprint rack types.
The operator places a rack fully or partially filled with capped tubes onto the instrument’s stage. The IntelliXcap confirms that the consumables match the expected height and then de- caps or caps all the tubes. If the instrument detects that the tube rack is of a height different to that expected, it returns an error-message.
Page 30
Ensure that the door is free from obstructions, and that there is nothing in front of the instrument. Turn the IntelliXcap on, using the switch at the rear of the system. The IntelliXcap will start up and begin the Initialization-process.
Page 31
Choose the Start button. The instrument confirms the height of the tubes and then starts de- capping. When in use, the IntelliXcap displays the current process with a large STOP button display indicating the unit is in use. In case of need, you have the option to stop the current process by clicking the button on screen or by pressing the E-Stop button located at the top of the unit.
Page 32
Part Number: 319430 Rev. A Step Action Once the de-capping process has finished, the IntelliXcap is ready to start the re-capping process. When ready, place the rack of uncapped tubes and press start. The instrument scans and detects the correct height of the tubes and begins the recapping process.
However, it is recommended that the preventative maintenance procedures and schedule provided in this section be followed to extend the operating life of the IntelliXcap. If additional procedures are required, they will be supplied along with their maintenance schedules by Brooks Automation.
"Technical Support" on page 1. Schedules and Procedures The service life of the IntelliXcap is 20 years of a daily 8-hour operation. This is based on the presumption that all service and maintenance instructions described in this instruction manual are observed.
Shield and/or keep the work area in a moist condition to prevent dust from flying around or smoldering. The operator, or specially trained cleaning staff, should tidy up and clean the IntelliXcap and its surroundings daily. During this work, the same requirements for the use of tools and personal protective equipment apply as for the operational work.
If necessary, use isopropyl alcohol to disinfect and further clean surfaces Inspect Cartridge The cartridges used on the IntelliXcap have a limited life, generally expected to be in the region of 5000 cycles. After every 2500 cycles with a cartridge, the IntelliXcap will give a service counter warning.
The system will also count the cycles for each cartridge type. If the user operates the IntelliXcap with two different kinds of cartridges, it counts and records the cycles separately for each type. The Cartridge warning can be removed by pressing the Reset button.
7. Troubleshooting IntelliXcap Technical Support Part Number: 319430 Rev. A Technical Support For technical support in connection with your IntelliXcap, please contact a technical representative at: BLSS.Service@brooks.com Location GUTS® Contact Number Website +1-800-FOR-GUTS (1-800-367-4887) North America http://www.brooks.com/ +1-978-262-2900 Europe +49-1804-CALL-GUTS (+49-1804-2255-4887) http://www.brooks.com/...
Manually add a new cap to the ERROR tubes and perform a new de- If the IntelliXcap fails on the second attempt, an error capping cycle. message is shown on the screen, and the IntelliXcap stops.
"Technical Support" on page 39. Additionally, it is possible to run the IntelliXcap units in a legacy mode, emulating the commands used on the older XSD96,48 and 24 units – for support with configuration and commands please contact Brooks support as described above.
"Technical Support" on page 39. Table 7-2: IntelliXcap 96 Part IntelliCartridges for IntelliXcap 96 Description – (single cartridge) Number 48-8013- 96 format cap driver cartridge with 96 individual cap drivers for FluidX Internal Next-Gen Thread 48-8013-...
Page 50
Error Recovery Part Number: 319430 Rev. A Table 7-3: IntelliXcap 48 Part IntelliCartridges for IntelliXcap 48 Description – (single cartridge) Number 48-8015- 48 format cap driver cartridge with 48 individual cap drivers for FluidX 1.5ml, 1.9ml, 4.0ml Next-Gen Jacket & 2.0ml Jacket Auto-friendly External Thread 48-8015- 48 format cap driver cartridge with 48 individual cap drivers for FluidX Non-Jacket 2.0ml /...