10) Press "Set" to upload the settings into the UV222.
Once this information is saved and programmed into the unit, it is ready to test/start up.
The unit will calculate the radiation automatically to stay within the limits of the ACGIH like
the example below:
Pre
disinfe.
If during the duty cycle a person comes closer to the unit than the programmed distance
from the eye, UV222 will automatically stop, and it will resume the activity 10 sec. after this
security distance is cleared.
If movement in the room is detected during the pre or post disinfection cycle, the unit will
stop working, and will resume its activity after no more movement is detected.
UV222
Technical & Installation Manual
18
Occupied timeslot
Time
A
Post
disinfe.
A
TLV
Need help?
Do you have a question about the UV222 and is the answer not in the manual?