Safety Information Indications for Use The CS 3800 is a digital optical scanning device used to record the topographic characteristics of teeth or dental impressions in three dimensions. The resulting topographic impressions are intended for use in the computer-aided design and manufacturing of dental restorative prosthetic devices, dental implant prosthetic devices, and orthodontic models.
Conventions in This Guide The following special messages emphasize information or indicate potential risks to personnel or equipment. WARNING: Warns you to avoid injury to yourself or others by following the safety instructions precisely. Caution: Alerts you to a condition that might cause serious damage.
This unit is not intended for use with flammable anesthetics or flammable agents. DO NOT pull or twist the cable. DO NOT drop the scanner or the accessories. CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01...
Page 8
1.83 meters away from the patient. If the equipment is faulty, turn it OFF, display an “Out of Service” notice, and contact a qualified Carestream Dental service technician. Using components, accessories, cables, and spare parts other than...
Page 9
Do not use any cell or battery which is not designed for use with the equipment. Always purchase the battery recommended by the device manufacturer for the equipment. CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01...
Page 10
Keep cells and batteries clean and dry. Wipe the cell or battery terminals with a clean, dry cloth if they become dirty. Do not leave a battery on prolonged charge when not in use. After extended periods of storage, it may be necessary to charge and discharge the cells or batteries several times to obtain maximum performance.
Remove the blood and/or body fluids with the dampened lint-free cloth. Caution: DO NOT touch the contact points on the battery or the scanner. Do not allow liquid to infiltrate the cover. CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01...
Disinfecting the Scanner After each patient, the scanner must be thoroughly disinfected. To adequately disinfect the scanner, follow the disinfectant manufacturer’s instructions for the appropriate contact time. Important: If the scanner is visibly soiled, it must be thoroughly cleaned prior to disinfecting. “Cleaning the Scanner.”...
Do not soak the scanner tips in disinfectant overnight. Dry the scanner tips thoroughly before mounting onto the scanner. Do not use an ultrasonic cleaning machine to clean the scanner tips. CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01...
Manually Cleaning the Scanner Tips To manually clean the scanner tips, follow these steps: Rinse excess soil from the tip. Using a soft brush, apply an enzymatic detergent solution (e.g., Metrex EmPower) to all surfaces. Rinse under clean, running water. Inspect the tip.
Minimum 3 Minutes 20-30 Minutes Gravity Autoclave Exposure Time at 132°C Exposure Time at 134°C Minimum Drying Time Minimum 15 Minutes Minimum 10 Minutes 15-30 Minutes Important: Do NOT exceed 134°C. CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01 11...
Precautions Before Use Perform the following activities on your product and accessories before use. Cleaning, Disinfecting, and Sterilizing To ensure maximum hygienic safety for the patient and to minimize the risk of cross-contamination, carefully perform the following maintenance activities on your scanner and accessories.
Manufacturer’s address Manufactured date CAUTION: Consult accompanying documentation. Refer to instruction manual/booklet. Direct current Medical device Name of the European authorized representative and address of the registered place of business. CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01 13...
Standard: Usability EN/IEC 62366-1 Medical Devices – Part 1: Application of usability engineering to medical devices EN 62304 / IEC 62304 Medical device software - Software life cycle Processes CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01 17...
Page 22
Compliance with European and International Standards EN ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process EN ISO 14971 Medical devices - Application of risk management to medical devices EN ISO 15223-1 Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General...
Page 23
Mode of operation Continuous operation Flammable anesthetics Not suitable for use in the presence of flammable anesthetics or a mixture of flammable anesthetics with air or oxygen or nitrous oxide. CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01...
WiFi The wireless version of the CS 3800 intraoral scanner operates with the 802.11ac protocol. Only channel 42 is used for the handpiece. The central frequency is 5210MHz. The channel bandwidth is 80MHz. The radio output power is 18.5 dBm (nominal).
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (IEC 60601-1-2) The CS 3800 is intended for use in the electromagnetic environment specified below. The customer or user of the CS 3800 should assure that it is used in such an environment. Emissions...
Page 26
Fully Compliant with IEC 60601-1-2: 2014 The CS 3800 is intended for use in the electromagnetic environment specified below. The customer or the user of the CS 3800 should assure that it is used in such an environment. Immunity Test IEC 60601 Test...
Page 27
Electromagnetic Immunity (IEC 60601-1-2) The CS 3800 is intended for use in the electromagnetic environment specified below. The customer or the user of the CS 3800 should assure that it is used in such an environment. Immunity IEC 60601 Test Level...
For the immunity to proximity fields from RF wireless communications equipment, the CS 3800 is compliant with the test levels specified below, according to IEC60601-1-2 standard. The customer or user of the CS 3800 should assure that it is used in such an environment. Test Frequency...
Sensor technology 1/2.9 inch CMOS Illumination LED: Red, Blue, Green Field of view 16 x 14 mm 13 x 7 mm (posterior tip) Depth of field -2 to +16 mm CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01 25...
Page 30
Anti-fogging By airflow technology Digital connection WiFi 5GHz, 802.11ac Handpiece 226 x 38 x 60 mm dimensions with normal tip or side tip Components Technical Specifications Weight 240 g (with battery) Handpiece Input: 5 V 5A (with backup power cable) 4.2 V 3A (with battery) Handpiece...
The All-In-One computer system requirement is an exception. The computer and its screen should be situated in or close to the operating area, in the visual field of the practitioner when using the CS 3800. Important: It is MANDATORY to check that your system configuration is compatible with the computer system requirements for the CS 3800 software.
Authorized Representative in Brazil Carestream Dental Brasil Eireli Rua Romualdo Davoli, 65 1º Andar, Sala 01 - São José dos Campos São Paulo - Brazil CEP (Zip code): 12238-577 CS 3800 Safety, Regulatory, and Technical Specifications User Guide (TA3977)_Ed01 29...