Page 2
The CS 1200 is intended for professional use only. US Federal law restricts this device to sale by or on the order of a dentist. Manual Name: CS 1200 Intraoral Camera User and Installation Guide Part Number: 9H1809 Revision Number: 05...
CAUTION: Alerts you to a condition that might cause serious damage. Important: Alerts you to a condition that might cause problems. Note: Emphasizes important information. Tip: Provides extra information and hints. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05...
“Packaging of the CS 1200 Camera” for the packaging components list. Inspect the packages of the CS 1200 intraoral camera after the shipment is received. If there are any signs of damage or missing parts, contact the shipping company and Carestream Health, Inc., immediately.
Page 8
Table 1 Packaging of the CS 1200 Camera Component Description Quantity DC power cable and adapters (AV/S-Video ONLY) Protective sheaths Document and software package: CS Intraoral Cameras Installation Wizard DVD Documents on DVD: • User and Installation Guide • Safety, Regulatory, and Technical Specifications User Guide Safety Information Sheet •...
CS 1200 Intraoral Camera Overview The CS 1200 intraoral camera is an easy-to-use, simple-to-share, and high-quality imaging system. It has three connection configurations based on the model you purchased: • • S-Video • When connected to a computer, it has the following functions: •...
Page 10
5 Cable connection • • S-Video • 6 Camera lens 7 Push-pull connector 8 USB connector 9 Camera holder 10 Protective sheaths (not pictured) Chapter 3 CS 1200 Intraoral Camera Overview...
• CS 1200 Acquisition interface CS 1200 Acquisition Interface Overview The CS 1200 Acquisition interface can acquire, review, and store videos and images of the oral cavity. Figure 2 Acquisition Interface Overview Title Bar: Displays the patient’s name. Toolbar: See “Toolbar Overview”...
Select a format dialog box to select a new display format. Mirror image and Video button: Mirrors the live video and the captured images. Full screen button: Displays the live video, recorded video, and the captured images in full screen. Chapter 4 CS 1200 Software Overview...
Transfer to Hard Disk button: Transfers images from the camera to the computer. Select Patient Dialog Box The Select Patient dialog box enables you to transfer images of the selected patient number to the Acquisition interface. Figure 3 Select Patient Dialog Box CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05...
The Tooth Chart dialog box enables you to select tooth numbers associated with the image or video. Figure 5 Tooth Chart Date field: Displays the current date. Comments field: Enables you to enter comments to be associated with an image or video. 10 Chapter 4 CS 1200 Software Overview...
Selections field: Enables you to select the display format you want. Preview field: Enables you to preview the selected display format. Cancel button: Cancels the format selection. Open button: Opens the format selection. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 11...
Capture sound: Enable/disable the audio sound when an image is captured. • Record sound: Enable/disable the audio sound when a video is recorded. Important: Ensure that the proper frequency setting is selected for your region. The wrong setting can negatively impact image quality. 12 Chapter 4 CS 1200 Software Overview...
Contrast Setting: Select the video contrast setting. Tools Preferences Figure 9 Tools Preferences The Tools Preferences dialog box enables you to select the following option: • Create log file: Save log files for service information. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 13...
Insert the CS Intraoral Cameras Installation Wizard DVD in the DVD drive. If the DVD does not run automatically, navigate to your DVD drive in Windows Explorer and double-click the Launcher.exe file. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 15...
Connecting the Camera to a Monitor To connect the camera to a monitor, follow the instructions in the figures below. S-Video Connection AC/DC AV Connection AC/DC Important: Isolate mains power by unplugging the power adapter. 16 Chapter 5 Setting Up the CS 1200 Intraoral Camera...
Remove the adhesive tape protection from the support (A). Place the taped side of the camera holder on the clean surface (B) and press firmly several times for correct adherence. The maximum adherence is obtained after two hours. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 17...
Page 22
Preferences” on page For places on which you cannot use the adhesive tape, insert a screw through the appropriate hole in the camera holder to affix it to a solid surface. 18 Chapter 5 Setting Up the CS 1200 Intraoral Camera...
Imaging window: In the • If you installed only the CS 1200 or you have installed the CS 1500 and/or CS 1600 camera before the CS 1200, click • If you have installed a K 1000 camera before the CS 1200, click the arrow button on , and select the CS 1200 camera from the drop-down list.
The Select Source window is displayed. Select CS Intraoral Camera 2.3 (32-32) and click Select. Open the Acquisition interface using the TWAIN interface; see the user guide for your imaging software. Note: The software used for image acquisition must be compliant with TWAIN.
Adjust the protective sheath if necessary. The end should be tight to prevent fogging or blurring of the image. Note: Installing the sheath incorrectly has a negative effect on image quality. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 21...
Acquiring Live Video or Images Accessing the Acquisition Interface To access the Acquisition interface: • If you use CS Imaging Software, click in the Imaging window and select your camera. • If you use imaging software via a TWAIN interface, see the user guide of your imaging software for details on opening the Acquisition interface.
“Tooth Chart Dialog Box” on page 10). The selected images are displayed and saved temporarily in the image gallery. Use the toolbar to view the image (see “Image Toolbar” on page CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 23...
Page 28
Press the Review button on the camera to review the images. Use the image gallery toolbar to scroll through, save, or delete the images (see “Image Gallery Toolbar” on page Click to close the Acquisition interface. The images are saved and displayed in the Patient history. Remove and dispose of the protective sheath.
You can have a total of nine patients listed in the dialog box at one time. Figure 12 Patient Number Dialog Box CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 25...
Preparing for Acquisition Preparing the Camera The protective sheath covers the camera and provides a sanitary shield for the patient. Use a new protective sheath with each patient. Always discard the protective sheath after each use. For additional protective sheaths, contact your dealer or the protective sheaths manufacturer. Important: Use only the protective sheaths designed specifically for this camera.
For details on how to delete images, see“Deleting Images” on page 29. For details on how to transfer images to the computer, see “Transferring Images” on page Note: The camera can store up to 300 images. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 27...
Remove and dispose of the protective sheath when you have finished with the current patient. Note: The camera automatically goes into sleep mode when inserted into the camera holder or left idle for 1.5 minutes. To use it again, pick it up or press the Capture button.
Software” on page 15. Press the ON/OFF button 3 seconds to turn on the camera. In the Imaging window, click and select your camera. The Acquisition interface is displayed. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 29...
On the Acquisition interface, click . The Select Patient dialog box is displayed. Select the patient number and click Transfer. When the transfer is complete, the images are deleted automatically from the camera. Transferring Images to the Computer Hard Disk To transfer images from the camera to the computer hard disk, follow these steps: Press the ON/OFF button 3 seconds to turn off the camera.
The camera is not delivered in disinfected condition. To ensure maximum hygienic safety for the patient, carefully follow the instructions to disinfect the camera before use. Important: For information on cleaning and disinfecting, see the Safety, Regulatory, and Technical Specifications User Guide. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 31...
Live video/recorded video is choppy or Make sure that the graphics card is using the latest lagging. drivers. Visit the website of the graphics card manufacturer to learn how to update the drivers. CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed05 33...
Page 38
Problem Description Action Click Start-->Control Panel (View by Large • Icons)-->Performance Information and Tools, Glitches with the display in Windows Vista and select Re-run the assessment to refresh the or 7. For example, a "picture in picture" rating score. effect in live video or a flashing screen Make sure that the graphics card is using the •...