Camera is powered up and on but green light is flashing and cannot connect. I have restarted the computer and USB outlet is ok
1 comments:
Mr. Anderson
April 10, 2025
The green light flashing on the Carestream DENTAL CS 1500 camera indicates that it is not connected or not communicating properly with the workstation. To connect it:
1. Use the proprietary USB cable that came with the camera. 2. If it's a wired camera, connect it directly to the computer using the rapid USB cable. 3. If it's a wireless camera, ensure it is paired with the dock station, which acts as the communication hub. 4. Make sure the CS Acquisition Interface is installed (check via Windows Control Panel). 5. In the CS Imaging Software, go to Options -> Preferences -> Camera, and make sure the video capture source box is checked. 6. Reinstall the acquisition software if needed.
This should establish communication and stop the green light from flashing.
Page 2
The information in this document is subject to change. Neither Carestream Dental LLC nor any of its subsidiaries shall be liable for errors contained herein or for incidental damages in conjunction with the furnishing, performance, or use of this material.
CAUTION: Alerts you to a condition that might cause serious damage. Important: Alerts you to a condition that might cause problems. Note: Emphasizes important information. Tip: Provides extra information and hints. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07...
“Packaging of the CS 1500 Camera” for the packaging components list. Inspect the packages of the CS 1500 intraoral camera after the shipment is received. If there are any signs of damage or missing parts, contact the shipping company and Carestream Dental LLC immediately.
Page 8
Table 1 Packaging of the CS 1500 Camera Description Quantity Rapid USB cable One additional USB cable with two USB connections. It is used only if the traditional USB cable with one USB connection cannot provide enough power to the camera.
CS 1500 Intraoral Camera Overview The CS 1500 intraoral camera is a high-quality dental imaging system, featuring 1mm-to-infinity auto-focus range and liquid lens technology. The camera brings a new level of usability to dental professionals to plan treatment and guide patients to more informed treatment decisions.
The camera reverts to sleep mode when it is in the holder. If the Note: camera is in the holder for more than 3 hours (default), it shuts down. To change the default automatic power-off time, see “General Preferences” on page Chapter 3 CS 1500 Intraoral Camera Overview...
Wired USB Intraoral Camera with Firmware 1.1.5.X / 1.0.5.X or above. Figure 3 Rapid USB Cable Overview Battery Battery indicator indicates the battery is charged. indicates battery charging is in progress. indicates the battery is not being charged. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07...
Page 12
Chapter 3 CS 1500 Intraoral Camera Overview...
CS 1500 Software Overview Minimum Computer System Requirements For the minimum computer system requirements, see the CS 1500 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide. Important: It is MANDATORY to check that your computer system configuration is compatible with the computer system requirements for the CS 1500.
The Acquisition interface is a user-friendly working interface that was designed and developed specifically for the CS 1500. It can acquire and review videos and images of the oral cavity and export them to the CS Imaging Software for storage.
If the icon is red, the USB is disconnected. Preference button: Accesses the Preferences dialog box. Information button: Accesses information on the software version, firmware version, and camera serial number. Online help button: Accesses the online help. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 11...
Brightness button: Adjusts the brightness of the image. Contrast button: Adjusts the contrast of the image. Scroll button: Increases or decreases the brightness and contrast of the image. Reset button: Clears all brightness and contrast changes made to the image. 12 Chapter 4 CS 1500 Software Overview...
Capture sound: Enable/disable the audio sound when an image is captured. Record sound: Enable/disable the audio sound when a video is recorded. Important: Ensure that the proper frequency setting is selected for your region. The wrong setting can negatively impact image quality. 14 Chapter 4 CS 1500 Software Overview...
Brightness Setting: Select the video brightness setting. Contrast Setting: Select the video contrast setting. Tools Preferences Figure 9 Tools Preferences The Tools Preferences dialog box enables you to save log files for service information. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 15...
Single wired USB camera, connected to several computers (one computer at a time) using a USB configuration (see “Setting Up the Wired USB Camera” on page 21). USB / USB USB / USB USB / USB CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 17...
Installing the Software To install the software on the computer, follow these steps: Download the CS Imaging Software from the Carestream Dental website, and install the software. See the CS Imaging Software documentation: CS Imaging - Installation and Configuration Guide (SM998), for more information.
Access the Acquisition interface. See “Accessing the Acquisition Interface” on page 25. A message asks you to upgrade the firmware. Click Yes. The firmware upgrade starts and the following windows are displayed: CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 19...
Page 24
Make sure that the firmware upgrade is successful. Reopen the Acquisition interface to see a live video without the firmware upgrade notification. Open the Information panel, and make sure the firmware version has been updated. 20 Chapter 5 Setting Up the CS 1500 Intraoral Camera...
If you need to release the cable, press the Cable Release (A) button and gently pull the cable out. Make sure the cable is locked into place. Insert the USB end of the cable into the USB port of the computer. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 21...
Page 26
USB camera, use the additional dual USB cable with two connections that is provided. Press the ON/OFF button 3 seconds to power ON the camera. 3 sec. 22 Chapter 5 Setting Up the CS 1500 Intraoral Camera...
Check the adherence regularly to ensure that it is secure. Important: Carestream Dental LLC is not responsible for damage involving the camera falling due to loosening of the adhesive tape. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 23...
Page 28
Note: The camera reverts to sleep mode when it is in the holder. If the camera is in the holder for more than 3 hours, it shuts down. To change the default automatic power-off time, see “General Preferences” on page 24 Chapter 5 Setting Up the CS 1500 Intraoral Camera...
To access the Acquisition interface, follow these steps: In the Imaging window, If you installed only the CS 1500, or you have installed the CS 1200 and/or CS 1600 camera before the CS 1500, click If you have installed a K 1000 camera before the CS 1500, click the arrow button on , and select the CS 1500 camera option from the drop-down list.
Using Imaging Software with TWAIN Starting the Imaging Software To acquire an image using imaging software via TWAIN, follow these steps: From your imaging software, such as Photoshop, open the TWAIN interface. For example, in Photoshop, go to Import > Select Source. The Select Source window is displayed.
Adjust the protective sheath if necessary. The end should be tight to prevent fogging or blurring of the image. Note: Installing the sheath incorrectly has a negative effect on image quality. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed06 27...
Recording Video and Capturing Images on a Computer Recording and Viewing Videos To record or view a video, follow these steps: Hold the camera over the area you want to view. Live video is displayed on the Preview screen. To record the video, on the Acquisition interface, click to record, and click again to stop the recording.
“Tooth Chart Dialog Box” on page 13). The selected images are displayed and saved temporarily in the image gallery. Use the toolbar to view the image (see “Image Toolbar” on page 11). CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 29...
Page 34
Use the image gallery toolbar to scroll through or to delete the still images (see “Image Gallery Toolbar” on page 12). Press the Shutter Release button halfway down and release it to return to the Preview window. Click to close the Acquisition interface. The recorded video is saved and displayed in the Patient history.
Maintenance Cleaning and Disinfecting the Camera You must clean and disinfect the CS 1500 and accessories regularly. The camera is not delivered in disinfected condition. To ensure maximum hygienic safety for the patient, carefully follow the instructions to disinfect the camera before use.
Make sure you use the USB cable provided with the CS 1500 intraoral camera. Specify the integrated graphic card for the CS 1500 Acquisition software as follows: Open NVIDIA Control Panel (right-click on the Desktop and select NVIDIA Control Panel).
Page 38
Problem Description Action The colors of the live video or On the Acquisition interface, click and select captured images are not accurate. Video Adjustment Preferences, then adjust contrast and brightness. Check the contrast and brightness of the computer monitor.
Troubleshooting the System Setup Error Message: ffdshow compatibility manager TW.exe Select Use ffdshow (always) and Don’t ask me again. Click OK. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 35...
Live Video Does Not Work Make sure the hardware acceleration of the computer graphics card is correctly set up. To set the hardware acceleration of the computer graphics card, follow these steps: Right-click the desktop window and select Properties. Click the Settings tab. Click Advanced.
Page 41
Click the Troubleshoot tab. Set Hardware acceleration to FULL. Select Enable write combining. Click Apply. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 37...
List of Importers for European Union According to the MDR 2017/745 Carestream Dental France SAS 4 Rue F. Pelloutier, Croissy-Beaubourg 77435 Marne-la-Vallée Cedex 2, France Carestream Dental Germany GmbH Hedelfinger Str. 60 70327 Stuttgart, Germany CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 39...
Carestream Dental Spain, S.L.U. Paseo de la Castellana, 79 Madrid 28046, España Carestream Dental Italy S.r.l. Via Mario Idiojmi 3/3 Assago 20090 (MI), Italia List of Importers for Switzerland CURADEN AG Riedstrasse 12 CH-8953 Dietikon Switzerland Dema dent AG Furtbachstrasse 16 CH-8107 Buchs Switzerland...
Need help?
Do you have a question about the CS 1500 and is the answer not in the manual?
Questions and answers
Camera is powered up and on but green light is flashing and cannot connect. I have restarted the computer and USB outlet is ok
The green light flashing on the Carestream DENTAL CS 1500 camera indicates that it is not connected or not communicating properly with the workstation. To connect it:
1. Use the proprietary USB cable that came with the camera.
2. If it's a wired camera, connect it directly to the computer using the rapid USB cable.
3. If it's a wireless camera, ensure it is paired with the dock station, which acts as the communication hub.
4. Make sure the CS Acquisition Interface is installed (check via Windows Control Panel).
5. In the CS Imaging Software, go to Options -> Preferences -> Camera, and make sure the video capture source box is checked.
6. Reinstall the acquisition software if needed.
This should establish communication and stop the green light from flashing.
This answer is automatically generated