Carestream DENTAL CS 1500 User Manual
Hide thumbs Also See for CS 1500:

Advertisement

User's Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CS 1500 and is the answer not in the manual?

Questions and answers

Stacy
April 10, 2025

Camera is powered up and on but green light is flashing and cannot connect. I have restarted the computer and USB outlet is ok

1 comments:
Mr. Anderson
April 10, 2025

The green light flashing on the Carestream DENTAL CS 1500 camera indicates that it is not connected or not communicating properly with the workstation. To connect it:

1. Use the proprietary USB cable that came with the camera.
2. If it's a wired camera, connect it directly to the computer using the rapid USB cable.
3. If it's a wireless camera, ensure it is paired with the dock station, which acts as the communication hub.
4. Make sure the CS Acquisition Interface is installed (check via Windows Control Panel).
5. In the CS Imaging Software, go to Options -> Preferences -> Camera, and make sure the video capture source box is checked.
6. Reinstall the acquisition software if needed.

This should establish communication and stop the green light from flashing.

This answer is automatically generated

Subscribe to Our Youtube Channel

Summary of Contents for Carestream DENTAL CS 1500

  • Page 1 User’s Guide...
  • Page 2 The information in this document is subject to change. Neither Carestream Dental LLC nor any of its subsidiaries shall be liable for errors contained herein or for incidental damages in conjunction with the furnishing, performance, or use of this material.
  • Page 3: Table Of Contents

    Guide Chapter 2 CS 1500 System Description ....3 CS 1500 System Packaging of the CS 1500 Camera ....3 Packaging Chapter 3...
  • Page 4 Maintenance Chapter 9 Troubleshooting a CS 1500 Wired USB Intraoral Troubleshooting Camera......33 Troubleshooting the System Setup .
  • Page 5: Conventions In This Guide

    CAUTION: Alerts you to a condition that might cause serious damage. Important: Alerts you to a condition that might cause problems. Note: Emphasizes important information. Tip: Provides extra information and hints. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07...
  • Page 6 Chapter 1 Conventions in This Guide...
  • Page 7: Cs 1500 System Description

    “Packaging of the CS 1500 Camera” for the packaging components list. Inspect the packages of the CS 1500 intraoral camera after the shipment is received. If there are any signs of damage or missing parts, contact the shipping company and Carestream Dental LLC immediately.
  • Page 8 Table 1 Packaging of the CS 1500 Camera Description Quantity Rapid USB cable One additional USB cable with two USB connections. It is used only if the traditional USB cable with one USB connection cannot provide enough power to the camera.
  • Page 9: Camera Overview

    CS 1500 Intraoral Camera Overview The CS 1500 intraoral camera is a high-quality dental imaging system, featuring 1mm-to-infinity auto-focus range and liquid lens technology. The camera brings a new level of usability to dental professionals to plan treatment and guide patients to more informed treatment decisions.
  • Page 10: Camera Holder Components Overview

    The camera reverts to sleep mode when it is in the holder. If the Note: camera is in the holder for more than 3 hours (default), it shuts down. To change the default automatic power-off time, see “General Preferences” on page Chapter 3 CS 1500 Intraoral Camera Overview...
  • Page 11: Rapid Usb Cable Overview

    Wired USB Intraoral Camera with Firmware 1.1.5.X / 1.0.5.X or above. Figure 3 Rapid USB Cable Overview Battery Battery indicator  indicates the battery is charged.  indicates battery charging is in progress.  indicates the battery is not being charged. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07...
  • Page 12 Chapter 3 CS 1500 Intraoral Camera Overview...
  • Page 13: Minimum Computer System Requirements

    CS 1500 Software Overview Minimum Computer System Requirements For the minimum computer system requirements, see the CS 1500 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide. Important: It is MANDATORY to check that your computer system configuration is compatible with the computer system requirements for the CS 1500.
  • Page 14: Acquisition Interface Overview

    The Acquisition interface is a user-friendly working interface that was designed and developed specifically for the CS 1500. It can acquire and review videos and images of the oral cavity and export them to the CS Imaging Software for storage.
  • Page 15: Image Toolbar

    If the icon is red, the USB is disconnected. Preference button: Accesses the Preferences dialog box. Information button: Accesses information on the software version, firmware version, and camera serial number. Online help button: Accesses the online help. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 11...
  • Page 16: Image Gallery Toolbar

    Brightness button: Adjusts the brightness of the image. Contrast button: Adjusts the contrast of the image. Scroll button: Increases or decreases the brightness and contrast of the image. Reset button: Clears all brightness and contrast changes made to the image. 12 Chapter 4 CS 1500 Software Overview...
  • Page 17: Tooth Chart Dialog Box

    Adult dental arch: Enables tooth selection for an adult. Child dental arch: Enables tooth selection for a child. Reset button: Clears all tooth selections. Cancel button: Cancels all tooth selections. Save button: Saves tooth selections. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 13...
  • Page 18: Preferences Dialog Boxes Overview

    Capture sound: Enable/disable the audio sound when an image is captured.  Record sound: Enable/disable the audio sound when a video is recorded. Important: Ensure that the proper frequency setting is selected for your region. The wrong setting can negatively impact image quality. 14 Chapter 4 CS 1500 Software Overview...
  • Page 19: Video Adjustment Preferences

    Brightness Setting: Select the video brightness setting.  Contrast Setting: Select the video contrast setting. Tools Preferences Figure 9 Tools Preferences The Tools Preferences dialog box enables you to save log files for service information. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 15...
  • Page 20 16 Chapter 4 CS 1500 Software Overview...
  • Page 21: Cs 1500 Camera Configuration Options

    Single wired USB camera, connected to several computers (one computer at a time) using a USB configuration (see “Setting Up the Wired USB Camera” on page 21). USB / USB USB / USB USB / USB CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 17...
  • Page 22: Installing The Software

    Installing the Software To install the software on the computer, follow these steps: Download the CS Imaging Software from the Carestream Dental website, and install the software. See the CS Imaging Software documentation: CS Imaging - Installation and Configuration Guide (SM998), for more information.
  • Page 23: Upgrading The Firmware

    Access the Acquisition interface. See “Accessing the Acquisition Interface” on page 25. A message asks you to upgrade the firmware. Click Yes. The firmware upgrade starts and the following windows are displayed: CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 19...
  • Page 24 Make sure that the firmware upgrade is successful.  Reopen the Acquisition interface to see a live video without the firmware upgrade notification.  Open the Information panel, and make sure the firmware version has been updated. 20 Chapter 5 Setting Up the CS 1500 Intraoral Camera...
  • Page 25: Setting Up The Wired Usb Camera

    If you need to release the cable, press the Cable Release (A) button and gently pull the cable out. Make sure the cable is locked into place. Insert the USB end of the cable into the USB port of the computer. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 21...
  • Page 26 USB camera, use the additional dual USB cable with two connections that is provided. Press the ON/OFF button 3 seconds to power ON the camera. 3 sec. 22 Chapter 5 Setting Up the CS 1500 Intraoral Camera...
  • Page 27: Installing The Camera Holder

    Check the adherence regularly to ensure that it is secure. Important: Carestream Dental LLC is not responsible for damage involving the camera falling due to loosening of the adhesive tape. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 23...
  • Page 28 Note: The camera reverts to sleep mode when it is in the holder. If the camera is in the holder for more than 3 hours, it shuts down. To change the default automatic power-off time, see “General Preferences” on page 24 Chapter 5 Setting Up the CS 1500 Intraoral Camera...
  • Page 29: Accessing The Acquisition Interface

    To access the Acquisition interface, follow these steps: In the Imaging window,  If you installed only the CS 1500, or you have installed the CS 1200 and/or CS 1600 camera before the CS 1500, click  If you have installed a K 1000 camera before the CS 1500, click the arrow button on , and select the CS 1500 camera option from the drop-down list.
  • Page 30: Using Imaging Software With Twain

    Using Imaging Software with TWAIN Starting the Imaging Software To acquire an image using imaging software via TWAIN, follow these steps: From your imaging software, such as Photoshop, open the TWAIN interface. For example, in Photoshop, go to Import > Select Source. The Select Source window is displayed.
  • Page 31: Preparing For Acquisition

    Adjust the protective sheath if necessary. The end should be tight to prevent fogging or blurring of the image. Note: Installing the sheath incorrectly has a negative effect on image quality. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed06 27...
  • Page 32: Recording Video And Capturing Images On A Computer

    Recording Video and Capturing Images on a Computer Recording and Viewing Videos To record or view a video, follow these steps: Hold the camera over the area you want to view. Live video is displayed on the Preview screen. To record the video, on the Acquisition interface, click to record, and click again to stop the recording.
  • Page 33: Capturing And Viewing Images

    “Tooth Chart Dialog Box” on page 13). The selected images are displayed and saved temporarily in the image gallery. Use the toolbar to view the image (see “Image Toolbar” on page 11). CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 29...
  • Page 34 Use the image gallery toolbar to scroll through or to delete the still images (see “Image Gallery Toolbar” on page 12). Press the Shutter Release button halfway down and release it to return to the Preview window. Click to close the Acquisition interface. The recorded video is saved and displayed in the Patient history.
  • Page 35: Cleaning And Disinfecting The Camera

    Maintenance Cleaning and Disinfecting the Camera You must clean and disinfect the CS 1500 and accessories regularly. The camera is not delivered in disinfected condition. To ensure maximum hygienic safety for the patient, carefully follow the instructions to disinfect the camera before use.
  • Page 36 32 Chapter 8 Maintenance...
  • Page 37: Troubleshooting A Cs 1500 Wired Usb Intraoral Camera

    Make sure you use the USB cable provided with the CS 1500 intraoral camera.  Specify the integrated graphic card for the CS 1500 Acquisition software as follows: Open NVIDIA Control Panel (right-click on the Desktop and select NVIDIA Control Panel).
  • Page 38 Problem Description Action The colors of the live video or  On the Acquisition interface, click and select captured images are not accurate. Video Adjustment Preferences, then adjust contrast and brightness.  Check the contrast and brightness of the computer monitor.
  • Page 39: Troubleshooting The System Setup

    Troubleshooting the System Setup Error Message: ffdshow compatibility manager TW.exe Select Use ffdshow (always) and Don’t ask me again. Click OK. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 35...
  • Page 40: Live Video Does Not Work

    Live Video Does Not Work Make sure the hardware acceleration of the computer graphics card is correctly set up. To set the hardware acceleration of the computer graphics card, follow these steps: Right-click the desktop window and select Properties. Click the Settings tab. Click Advanced.
  • Page 41 Click the Troubleshoot tab. Set Hardware acceleration to FULL. Select Enable write combining. Click Apply. CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 37...
  • Page 42 38 Chapter 9 Troubleshooting...
  • Page 43: Manufacturer's Address

    List of Importers for European Union According to the MDR 2017/745 Carestream Dental France SAS 4 Rue F. Pelloutier, Croissy-Beaubourg 77435 Marne-la-Vallée Cedex 2, France Carestream Dental Germany GmbH Hedelfinger Str. 60 70327 Stuttgart, Germany CS 1500 Intraoral Camera User’s Guide (9H0951)_Ed07 39...
  • Page 44: List Of Importers For Switzerland

    Carestream Dental Spain, S.L.U. Paseo de la Castellana, 79 Madrid 28046, España Carestream Dental Italy S.r.l. Via Mario Idiojmi 3/3 Assago 20090 (MI), Italia List of Importers for Switzerland CURADEN AG Riedstrasse 12 CH-8953 Dietikon Switzerland Dema dent AG Furtbachstrasse 16 CH-8107 Buchs Switzerland...
  • Page 45 For more information, visit: www.carestreamdental.com ©Carestream Health, Inc., 2023. 2022-12 9H0951 Ed07...