Page 2
The CS 1200 is intended for professional use only. US Federal law restricts this device to sale by or on the order of a dentist. Manual Name: CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide Part Number: 9H1808...
WARNING: Warns you to avoid injury to yourself or others by following the safety instructions. Important: Alerts you to a condition that might cause problems. Note: Emphasizes important information. Tip: Provides extra information and hints. CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...
Warnings and Safety Instructions Warnings Camera • You MUST read and understand this safety information before using the camera. • The camera is not intended to be used with high frequency surgical equipment. • Before using the camera, check the outer surfaces of the camera and any accessories to ensure there are no rough surfaces, sharp edges, or protrusions which may cause a safety hazard.
Page 7
To power off the device, push the power button for 3 seconds. To isolate the device from mains supply, unplug the USB/AV/S-Video cables and the adapter from the power outlet. CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...
Computer • DO NOT place the computer and the peripheral equipment connected to it in the immediate vicinity of the patient. Leave at least 1.83 m distance between the patient and the equipment. • The camera is only intended to be connected to the computer or monitor which is at least IEC 60950 or equivalent standards certified.
Hygiene and Disinfection Perform the following maintenance activities on your CS 1200 and accessories regularly. To ensure maximum hygienic safety for the patient, carefully follow the instructions to prepare the CS 1200 for use. To minimize the risk for cross-contamination, after each patient, clean and disinfect the CS 1200.
Cleaning the Camera If the camera is visibly contaminated with blood and/or body fluids, you must clean the camera before disinfecting it. To clean the camera, follow these steps: Dampen (do not soak) a lint-free cloth with lukewarm water. Remove the blood and/or body fluids with the dampened lint-free cloth.
Visually inspect the camera for signs of deterioration, especially around the buttons and the cable. If damage is noted, do not use the camera and contact your representative. CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...
Contact your local sales representative for additional information on the collection and recovery programs available for this product. Manufacturer’s address Attention: Consult accompanying documentation. Refer to instruction manual/booklet. Direct current CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...
Indications for Use The CS 1200 intraoral camera is indicated for use by health professionals in viewing and capturing intraoral or extraoral color video images for the purposes of: • Allowing practitioners to view regions of the oral cavity. •...
Guidance and Manufacturer’s Declarations Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The CS 1200 intraoral camera is intended for use in the electromagnetic environment specified below. The customer or user of the intraoral camera should assure that it is used in such an environment.
Page 17
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The CS 1200 intraoral camera is intended for use in the electromagnetic environment specified below. The customer or the user of the intraoral camera should assure that it is used in such an environment.
Electromagnetic Immunity for Equipment and Systems Fully Compliant with IEC 60601-1-2: 2007 The CS 1200 intraoral camera is intended for use in the electromagnetic environment specified below. The customer or the user of the intraoral camera should assure that it is used in such an environment.
Recommended Separation Distance Between Portable and Mobile RF Communications Equipment and the Intraoral Camera System The CS 1200 intraoral camera is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the...
EMC Standards for Intraoral Camera IEC 60601-1-2: 2007 EMC requirements and tests, Medical Electrical Equipment including CISPR11:2003 +A1: 2004 + A2:2006 Group 1, Class B. NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
EN ISO 14971: Medical devices - Application of risk management to medical devices EN 980: Symbols for use in the labeling of medical devices EN 1041: Information supplied by the manufacturer of medical devices CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...
Other Equipment The CS 1200 intraoral camera should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the CS 1200 intraoral camera should be observed to verify normal operation in the configuration in which it will be used.
Video: 640 x 480 pixels Lighting White LEDs 80° Focus range 3 mm - 25 mm Connecting cable length Connection: Computer USB 2.0 high speed Connection: Monitor S-Video; AV CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...
Page 24
(Continued) Components Technical Specifications Output impedance of S-Video and AV 75 ohm cable Dimension 207 x 29x 22 mm Weight of camera without cable 60 g Adapter • Input: 100-240V ~ 50/60Hz, 400mA Adapter • Output: 5V 1.5A Environmental Requirements +5 ~ +30 °C Operating temperature Transportation and storage...
Note: The quality of images is affected by the quality of the monitor and monitor settings. See your monitor user's guide for information. CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...
Page 27
Representante no Brasil Carestream do Brasil Comércio e Serviços de Produtos MédicosLtda. Rua Pequetita, 215 cjs. 31 E 32 Edifício Atrium VII - Vila Olímpia São Paulo - Brazil CEP (Zip code): 04552-060 CS 1200 Intraoral Camera Safety, Regulatory, and Technical Specifications User Guide (9H1808)_Ed03...