The Cephalometric Modality User guide includes information on the Cephalometric modality for the CS 8100 and CS 8100 3D Families. For Panoramic or 3D modalities, see the Panoramic Modality User guide for the CS 8100 Family or the Panoramic and 3D Modality User guide for the CS 8100 3D Family.
CAUTION: Alerts you to a condition that might cause serious damage. Important: Alerts you to a condition that might cause problems. Note: Emphasizes important information. Tip: Provides extra information and hints. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 7
CS 8100SC Access: the panoramic modality (without the 2D+ radiological exam modality) and the cephalometric modality (without the 26x24 FoV (Field of View)). CS 8100 and CS 8100 Access can be upgraded to Cephalometric Modality, when the Scan Ceph module is provided as an upgrade kit.
Mobile Components Figure 1 illustrates the: Up and down movement of the unit mobile components Rotation and translation movement of the rotative arm and the cephalostat Figure 1 CS 8100SC Unit Mobile Components Figure 2 CS 8100SC 3D Unit Mobile Components The patient can enter either the left side or the right Important: side of the unit.
17 X-ray remote control Head and chin rest 18 Carpus panel (optional) Chin rest and bite block 19 Computer hosting the imaging and the acquisition software 10 Panoramic digital sensor Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 10
Figure 4 CS 8100SC 3D Functional Components Unit rotative arm 11 Unit column Unit head 12 Frankfort positioning indicator Emergency stop knob 13 Secondary collimator ON/OFF button 14 Nasion support Cephalostat arm 15 Head clamps and ear cones Cephalostat head 16 Cephalometric digital sensor X-ray source assembly 17 X-ray remote control...
The following figures illustrates the locations of the digital sensors and cepahometric sensors of the units. Figure 5 CS 8100SC digital sensor and cephalometric sensor locations Figure 6 CS 8100SC 3D digital sensor and cephalometric sensor locations Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Head and Chin Rest CS 8100 Family - Head and Chin Rest Figure 7 illustrates the functional components of the CS 8100 Family head and chin rest. Figure 7 CS 8100 Family Head and Chin Rest - Functional Components Positioning Panel...
CS 8100 3D Family - Head and Chin Rest Figure 8 illustrates the functional components of the CS 8100 3D Family head and chin rest. Figure 8 CS 8100 3D Family Head and Chin Rest - Functional Components Positioning Panel...
Ready Indicator LED: Green indicates that the unit is ready for acquisition. Frontal head rest adjustment knob (CS 8100 Family only): Positions the up or down inclination of the patient head by turning the knob.
The following accessories are used when positioning a patient using the Cephalometric Modality. Table 1 Cephalometric Modality Positioning Accessories Accessory Description Single use sheaths for ear cones Carpus panel (optional) Frankfort positioning indicator Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
X-ray Remote Control Overview The X-ray remote control enables you to launch a radiological image acquisition via the exposure button from outside the X-ray room. You must press and hold the exposure button until the end of acquisition. Premature release of the exposure button interrupts the acquisition.
Computer System Requirements For minimum computer system requirements and configuration for the CS Imaging Software, see the CS 8100 Family and CS 8100 3D Family Safety, Regulatory and Technical Specifications User Guides. Important: It is MANDATORY to check that the computer system configuration is compatible with the computer system requirements for the CS 8100 Family or CS 8100 3D Family software used by you.
Cephalometric Acquisition Interface Overview The Acquisition interface provides you with imaging acquisition functions. Figure 1 Cephalometric Acquisition Interface Information button: Show versions: Identifies Software and Firmware versions. Resetting Anatomy setting: Resets to the manufacturing parameter settings. Memorize Anatomy setting: Memorizes the user preference settings for each patient type (kV, mA and seconds).
Page 19
Click Patient to select patient type parameters. Click Parameters to select exposure parameter options. The Product name and logo is indicated here on the acquisition interface. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 20
Cephalometric Acquisition interface Program Pane The Cephalomentric Acquisition Interface Program Pane enables you to choose the different radiological exams and acquisition formats. Figure 2 Cephalometric Acquisition Interface Program Pane 1 Radiological exam options: Click for an oblique exam. Click for a lateral exam. Click for a submento-vertex exam.
Page 21
Important: See the CS 8100 Family and CS 8100 3D Family Safety, Regulatory and Technical Specifications User Guides for information on radiation protection and recommendations when you select a patient type, especially the pediatric patient. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 22
Cephalometric Acquisition Interface Parameter Pane The Cephalometric Acquisition Interface Parameter Pane enables you to choose exposure parameters for the radiological image acquisition. If the default parameter setting is not adapted to your patient type, you can manually adapt the parameter setting to the patient type and save this setting as the default setting.
Repeat Steps 2 to 4 for the following parameters: 80 kV - 10 mA 85 kV - 10 mA The unit is now ready to be used for acquisition. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Firewall or Network Settings You will not be able to access the Cephalometric Acquisition interface if you have not configured the firewall or network settings. To configure the firewall or network settings, follow these steps: On your desktop, double-click to open the CS Imaging Software. The Windows Security Alert dialog box displays.
Cephalometric Acquisition Interface. The Cephalometric Acquisition interface is displayed. Chapter 5—Acquiring Images using the Cephalometric Modality for information on how to launch an acquisition. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Program pane exam options. When you position the head clamps manually, the relevant exam option icon becomes active. the Cephalometric Acquisition Interface, select the appropriate FoV option. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 28
Child Adult: Small, Medium, Large Important: See the CS 8100 Family and CS 8100 3D Family Safety, Regulatory and Technical Specifications User Guides for information on radiation protection and recommendations when you select a patient type especially the pediatric patient.
Place both ears in between the head clamps. On the positioning panel, press and hold until you get the ear cones to the same height of ear auditory canals. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 30
Gently push in at the top part of a head clamp bar to retract both head clamp bars and to fit the ear cones into the ear auditory canals. Lower the nasion support to a vertical position. Use the Frankfort positioning indicator as a visual aid to adjust the up or down inclination of the patient head for a Frankfort plane horizontal alignment.
The radiation emission dose is expressed in mGy.cm . This dose is measured at the primary collimator outlet. The dose is accurate to +/-30%. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Acquiring a Frontal AP and PA Image for the Adult and Pediatric Patient Before acquiring an image, check that you have: Selected the patient record. Accessed the CS Imaging Software. Accessed the Cephalometric Acquisition Interface. Preparing the Unit and Setting the Acquisition Parameters To set the acquisition parameters, follow these steps: On the unit, position the head clamps manually for the frontal exam.
Page 33
(Optional) If the default parameter setting is not adapted to your patient type, click the Parameter button: Select the appropriate parameters. Click and select Memorize Anatomy setting if you want to save the new parameter settings as future default settings. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Preparing and Positioning the Adult and Pediatric Patient To prepare and position the patient, follow these steps: Ask the patient to remove all metal objects. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders. Pull out at the top part of a head clamp bar to extend both bars.
Page 35
Gently push in at the top part of a head clamp bar to retract both head clamp bars and to fit the ear cones into the ear auditory canals. Figure 5–1 Frontal AP Figure 5–2 Frontal PA Figure 5–3 Frontal AP Figure 5–4 Frontal PA Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Launching the X-ray To launch the X-ray, follow these steps: Leave the X-ray room and close the door. You must keep visual contact with the patient during acquisition. Important: If you have a problem that requires you to stop the acquisition, release the exposure button of the remote control or press the red emergency stop button.
The radiation emission dose is expressed in mGy.cm . This dose is measured at the primary collimator outlet. The dose is accurate to +/-30%. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Acquiring an Oblique Image for the Adult and Pediatric Patient Before acquiring an image, check that you have: Selected the patient record. Accessed the CS Imaging Software. Accessed the Cephalometric Acquisition Interface. Preparing the Unit and Setting the Acquisition Parameters To set the acquisition parameters, follow these steps: On the unit, position the head clamps manually for an oblique exam.
Ask the patient to do the following: Move forward. Sit or stand up straight. Place both ears in between the head clamps. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
On the control panel, press and hold to level the ear cones to the patient’s auditory canals. Gently push in at the top part of a head clamp bar to retract both head clamp bars and to fit the ear cones into the ear auditory canals. Lower the nasion support to a vertical position.
The radiation emission dose is expressed in mGy.cm . This dose is measured at the primary collimator outlet. The dose is accurate to +/-30%. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Acquiring a Submento-Vertex Image for the Adult and Pediatric Patient Before acquiring an image, check that you have: Selected the patient record. Accessed the CS Imaging Software. Accessed the Cephalometric Acquisition Interface. Preparing the Unit and Setting the Acquisition Parameters To set the acquisition parameters, follow these steps: On the unit, position the head clamps manually for a frontal AP exam.
Ask the patient to do the following: Move forward. Sit or stand up straight. Place both ears in between the head clamps. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 44
On the control panel, press and hold to level the ear cones to the patient’s auditory canals. Gently push in at the top part of a head clamp bar to retract both head clamp bars and to fit the ear cones into the ear auditory canals. 40 Chapter 5 Acquiring Images using the Cephalometric Modality...
Important: If you have a problem that requires you to stop the acquisition, release the exposure button of the remote control or press the red emergency stop button. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Launch the X-ray with the remote control. Press and hold the exposure button until the end of the acquisition is indicated by the on-screen “Release Handswitch” message. The turns yellow and a warning sound is heard, indicating X-ray emission. When the acquisition ends, the Acquisition interface disappears and the acquired image is automatically transferred to the CS Imaging Software.
When the carpus panel is connected to the unit, the acquisition interface is set automatically in the and the Cephalometric Acquisition Interface carpus exam option becomes active. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Child Adult: Small, Medium, Large Important: See the CS 8100 Family and CS 8100 3D Family Safety, Regulatory and Technical Specifications User Guides for information on radiation protection and recommendations when you select a patient type especially the pediatric patient.
X-ray emission range and extends the hand to the carpus panel. To launch the X-ray, follow these steps: Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Leave the X-ray room and close the door. You must keep visual contact with the patient during acquisition. Important: If you have a problem that requires you to stop the acquisition, release the exposure button of the remote control or press the red emergency stop button.
On your desktop, double-click . The Service Tools window is displayed. 2. In the left pane, double-click Image Quality Tool. The Image Quality Control window is displayed. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...
Page 52
3. Select the desired test and follow the on-screen instructions. 48 Chapter 6 Maintenance...
I 17 The sensor is in the positioning Wait for the end of the positioning process. process for either panoramic, 3D or Ceph mode. Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04...