Carestream DENTAL CS 8200 3D User Manual
Hide thumbs Also See for CS 8200 3D:
Table of Contents

Advertisement

User Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CS 8200 3D and is the answer not in the manual?

Questions and answers

Summary of Contents for Carestream DENTAL CS 8200 3D

  • Page 1 User Guide...
  • Page 2 Notice Congratulations on your purchase of the Panoramic and 3D Modality for the CS 8200 3D. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The Panoramic and 3D Modality User guide for the CS 8200 3D provides information on complete or segmented tomographic digital panoramic X-ray image and three-dimensional digital X-ray image features.
  • Page 3: Table Of Contents

    Setting the Acquisition Parameters ....27 Preparing and Positioning the Adult and Pediatric Patient . . 30 Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 4 Launching the X-ray..... . 32 X-Ray Dose Emission Information ....33 Acquiring Sinus Images for the Adult and Pediatric Patient .
  • Page 5 MDR 2017/745 ......71 List of Importers for Switzerland ....72 Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 7: Conventions In This Guide

    CAUTION: Alerts you to a condition that might cause serious damage. Important: Alerts you to a condition that might cause problems. Note: Emphasizes important information. Tip: Provides extra information and hints. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 8 Chapter 1 Conventions in This Guide...
  • Page 9 CS 8200 3D Overview The following chart illustrates the different product configurations of the CS 8200 3D: CS 8200 3D Scan Ceph for CS 8200 3D Family 4x4 - 5x5 - 5x8 Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 10: Mobile Components

    CS 8200 3D Scan Ceph for CS 8200 3D Family 8x9 - 8x5 10x10 - 10x5 - 12x10 - 12x5 Upgradable Available WARNING: X-rays can be harmful and dangerous if not used properly. The instructions and warnings contained in this guide must be followed carefully.
  • Page 11 Figure 1 Unit Mobile Components Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 12: General Functional Components

    ON/OFF button Temple support Emergency stop knob X-ray source assembly Unit rotative arm Unit head Digital sensor X-ray remote control Unit column Computer hosting the imaging and the acquisition software Head and chin rest Chapter 2 CS 8200 3D Overview...
  • Page 13: Head And Chin Rest

    The following figure illustrates the functional components of the head and chin rest. Figure 3 Head and Chin Rest Functional Components Positioning Panel Bite block support Temple support adjustment knob Chin rest Temple support Hand grips Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 14: Positioning Accessories

    Frankfort guide bite block for panoramic (x3) Bite block for edentulous patients (x2) Single use sheaths for bite block (500 pcs box) Single use sheaths for 3D bite block and Frankfort guide bite block (100 pcs box) Chapter 2 CS 8200 3D Overview...
  • Page 15 (2) 3rd Molar offset 3D bite block (x2) (for use with 5x5, 4x4, 5x8 Ml and Mr FoV only) Temple support cone (for use with 3rd Molar offset 3D bite block only) Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 16: Positioning Panel

    Ready Indicator LED: Green indicates that the unit is ready for acquisition. Temple support adjustment knob: Opens and closes the temple supports when you turn the knob. 10 Chapter 2 CS 8200 3D Overview...
  • Page 17: X-Ray Remote Control Overview

    Exposure button: launches image acquisition. WARNING: X-rays can be harmful and dangerous if not used properly. The instructions and warnings contained in this guide must be followed carefully. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 18 12 Chapter 2 CS 8200 3D Overview...
  • Page 19: Computer System Requirements

    WARNING: It is MANDATORY to check that the computer system configuration is compatible with the computer system requirements for the CS 8200 3D software. If necessary you MUST update your computer system configuration. The CS 8200 3D MUST be connected to the computer via a point-to-point Ethernet link and not via a LAN.
  • Page 20: Acquisition Interface

    2a and 2b will appear in orange. Important: See the CS 8200 3D Safety, Regulatory and Technical Specifications User Guide (SMA67) for information on radiation protection and recommendations when you select a patient type, especially the pediatric patient.
  • Page 21 Menu: Provides access to the following sub-menus: General Settings, Reset to factory, About, Help. You can select the settings that you want as default settings for the acquisition interface. See “Displaying the User Preferences” Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 22: 3D Acquisition Interface Overview

    Important: See the CS 8200 3D Safety, Regulatory and Technical Specifications User Guide (SMA67) for information on radiation protection and recommendations when you select a patient type, especially the pediatric patient.
  • Page 23 13 Pop-up of accessory and patient head position: Will appear on the interface when you click to choose a radiological exam, to indicate which positioning accessory to use and how to position the patient head. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 24 14 Ready indicator:  Black indicates the unit is not ready to start acquisition.  Orange indicates the unit preparing for acquisition.  Green indicates the unit is ready to start acquisition. 15 Warning: Indicates that you must consult the accompanying document. 16 Ionizing radiation: ...
  • Page 25: Switching On The Unit

    Repeat Steps 2 to 4 for the following parameters:  80 kV - 10 mA  85 kV - 10 mA The unit is now ready to be used for acquisition. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 26: Firewall Or Network Settings

    Firewall or Network Settings You will not be able to access the Acquisition interface if you did not configured the firewall or network settings. To configured the firewall or network settings, follow these steps: On your desktop, double-click to open the CS Imaging Software. The Windows Security Alert dialog box displays.
  • Page 27: Accessing The Acquisition Interface

    Displaying the User Preferences To display the User Preferences window, follow these steps: 1. Click to open the Menu window. 2. Click General Settings. The User Preferences window displays: Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 28 3. Current UI language: Select the language of your choice and click Save. 3D Default acquisition program: Select the program of your choice and click Save. To select other preferences for your acquisition interface, click  Automatic MAR Reconstruction (optional): Activates the CS MAR filter. ...
  • Page 29 The balance of radiation exposure and image quality for the desired clinical task should be considered. You have the responsibility to determine the final settings of the device to achieve image quality. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02 23...
  • Page 30: Device Specific Features And Instructions

    Device Specific Features and Instructions The CS 8200 3D provides the following design features and instructions that enable safer use of our device with pediatric patients. Patient Size Selection The two smallest patient size icons resent the exposure values for children and adolescent patients.
  • Page 31: X-Ray Dose Information

    X-ray imaging outweighs the risks. The representative dose information associated with each radiological exam on the CS 8200 3D and the corresponding patient size is provided in the X-ray Dose Emission Information section of the CS 8200 3D Safety, Regulatory and Technical Specifications User Guide (SMA67).
  • Page 32 26 Chapter 5 Pediatric Use: Summary of Safety Instructions...
  • Page 33: Acquiring A Panoramic, Tmj X2, Tmj X4, Or Bitewing For The

    3. Select the radiological exam option for the X-ray image: Note: You can select the panoramic exams as full panoramic, segmented panoramic or reduced panoramic. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 34 When you select a radiological exam option, will display a pop-up illustration to show the positioning accessory that you must use and how to position the patient head. The pop-up will remain for three seconds and disappear. Click if you need to see it again. Click again to hide the pop-up.
  • Page 35 6. If the default parameter setting is not adapted to your patient type, click on the parameter display panel to open it and to select the appropriate parameters. To save the new parameter settings, click Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 36: Preparing And Positioning The Adult And Pediatric Patient

    Preparing and Positioning the Adult and Pediatric Patient To prepare and position the patient, follow these steps: 1. Ask the patient to remove and place all metal objects in the jewelry tray. 2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders.
  • Page 37  If you are using the standard panoramic bite block, ask the patient to bite into the groove of the bite block (B) as shown. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 38: Launching The X-Ray

     If you are using the Frankfort guide bite block for panoramic, ask the patient to bite on the bite block (B) as shown. Important: If you are using the Frankfort guide bite block, make sure the patient bites on the complete block, and that the patient’s occlusal plane inclination follows the angle of the bite block.
  • Page 39: X-Ray Dose Emission Information

    The radiation emission dose is expressed in mGy.cm . This dose is measured at the primary collimator outlet. The dose is accurate to +/-30%. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 40: Acquiring Sinus Images For The Adult And Pediatric Patient

     Child  Adult: Small, Medium, Large Important: See the CS 8200 3D Safety, Regulatory and Technical Specifications User Guide (SMA67) for information on radiation protection and recommendations when you select a patient type, especially the pediatric patient. 2. Click for sinus acquisition.
  • Page 41: Preparing And Positioning The Adult And Pediatric Patient

     Place chin on the sinus chin rest.  Position the feet slightly forward.  Relax and lower the shoulders for full motion of the unit rotative arm. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 42: Launching The X-Ray

    Note: C orrect posture reduces the shadow of the spinal column transferred to the image. 6. Ask the patient to do the following:  To close the eyes  To remain still  To breath through the nose  To not swallow Launching the X-ray To launch the X-ray, follow these steps: 1.
  • Page 43: X-Ray Dose Emission Information

    The radiation emission dose is expressed in mGy.cm . This dose is measured at the primary collimator outlet. The dose is accurate to +/-30%. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 44 38 Chapter 6 Acquiring Panoramic Images...
  • Page 45: Acquiring 3D Teeth Images For The Adult And Pediatric Patient

    2. Fit the 3D bite block support (A), and insert the appropriate 3D bite block (B). Make sure that you click the bite block well into place. Important: Use the 3D child bite block for a pediatric patient. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 46  If you are using a 3rd molar offset 3D bite block, insert the offset bite block (B) for a right-molar (M ) or left-molar (M ) acquisition as required: Important: The 3rd Molar offset 3D bite block is for use with 5x5, 4x4, and 5x8 Ml and MR FoVs only.
  • Page 47: Preparing And Positioning The Adult And Pediatric Patient

    3D child bite block for a pediatric patient. Cover the bite block with a hygienic barrier. 2. Ask the patient to remove and place all metal objects in the jewelry tray. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 48 3. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders. 4. Turn the adjustment knob to open the temple supports laterally. Note: You can adjust the temple supports manually to the back or front for a better head fit.
  • Page 49 8. Ask the patient to bite on the bite block:  If you are using a standard 3D bite block, ask the patient to bite on the 3D bite block as shown (A), (B). Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 50  If you are using a 3rd molar offset 3D bite block  To acquire an image of the right 3rd molar, ask the patient to bite on the side of the bite block indicated MR as shown (A):  To acquire an image of the left 3rd molar, ask the patient to bite on the side of the bite block indicated ML as shown (A): The indication on the bite block (ML or MR) should be placed in the corresponding region inside...
  • Page 51: Launching The X-Ray

    (which turned blue) and re-doing the scout view action.  Click to change the contrast from black to white or white to black. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 52: X-Ray Dose Emission Information

     To adjust for brightness or contrast, use  To set contrast change, brightness or contrast to default, click 3. Launch the X-ray with the remote control: Important: If you need to stop the acquisition due to a problem, release the exposure button of the remote control or press the red emergency stop button.
  • Page 53: Adult And Pediatric Patient

    Important: Use the dedicated 3D bite block to acquire 8x9, 8x5, 10x10, or 10x5 FoVs. Use the dedicated 3D bite block to acquire 12x10 or 12x5 FoVs. Make sure that you click the bite block well into place. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 54  Child  Adult: Small, Medium, Large Important: See the CS 8200 3D Safety, Regulatory and Technical Specifications User Guide (SMA67) for information on radiation protection and recommendations when you select a patient type, especially the pediatric patient. 6. If the default parameter setting is not adapted to your patient type, click on the parameter display panel to open it and to select the appropriate parameters.
  • Page 55: Preparing And Positioning The Adult And Pediatric Patient

    4. Ask the patient to enter the unit. On the Positioning panel, press and hold to raise the chin rest to the height of the patient. Note: If the patient is too tall, ask the patient to sit on a stool. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 56 5. Ask the patient to do the following:  Stand up straight (do not position the feet slightly forward).  Grip both the lower handles of the head and chin rest.  Relax and lower the shoulders for full motion of the unit rotative arm. 6.
  • Page 57: Launching The X-Ray

    If you want to perform a scout view  Click , then launch an X-ray with the remote control. The Scout view 2D screen will display the image: Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 58: X-Ray Dose Emission Information

     Once the scout view is done, the button becomes gray. You can perform another scout view by clicking the scout view button (which turned blue) and re-doing the scout view action.  Click to change the contrast from black to white or white to black. ...
  • Page 59: General Overview

    Slow Height Adjustment button: For slow adjustment of the height of the unit to the height of the object especially for the final adjustment. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02 53...
  • Page 60: Acquiring Images Of 3D Dental Objects

    Ready indicator:  Black indicates the unit is not ready to start acquisition.  Orange indicates the unit preparing for acquisition.  Green indicates the unit is ready to start acquisition. Warning: Indicates that you must consult the accompanying document. Ionizing radiation: ...
  • Page 61: Launching The Plaster Acquisition

    1. Select either the lower jaw, upper jaw, or full jaw acquisition mode. 2. See the on-screen illustration on how to correctly position the lower jaw, upper jaw, or full jaw plaster at the center of the black foam. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02 55...
  • Page 62: Launching The Plaster Acquisition With The Bite Index To Get The Occlusion

    3. Launch the X-ray with the remote control: Important: If you need to stop the acquisition due to a problem, release the exposure button of the remote control or press the red emergency stop button.  Press and hold the exposure button until the end of the acquisition is indicated by the on-screen “Release Switch”...
  • Page 63: Acquiring Images Of Impression Objects

    2. Place the dental impression holder on the 3D bite block support 3. Place the black foam on the dental impression holder Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02 57...
  • Page 64: Launching The Impression Acquisition

    4. Click Launching the Impression Acquisition To launch the acquisition, follow these steps: Important: Make sure that you position the acquisition object correctly to obtain quality data. 1. Select either the lower jaw or upper jaw acquisition mode. 2. See the on-screen illustration on how to correctly position the impression at the center of the black foam.
  • Page 65: Launching The Bite Index Scan To Register The Occlusion

    Note: You will be able to see the object in live-view mode on the workstation screen during the acquisition. When the acquisition ends, the acquired image appears on your screen. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02 59...
  • Page 66: Acquiring Images Of Triple Tray Impression Objects

    Acquiring Images of Triple Tray Impression Objects Preparing the Unit and Setting the Acquisition Options To launch the acquisition, follow these steps: 1. Insert the 3D bite block support in the chin rest base. 2. Place the dental impression holder on the 3D bite block support 3.
  • Page 67: Acquiring Images Of Appliance Objects

    When the acquisition ends, the acquired image appears on your screen. Acquiring Images of Appliance Objects Preparing the Unit and Setting the Acquisition Options To launch the acquisition, follow these steps: Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02 61...
  • Page 68 1. Insert the 3D bite block support in the chin rest base. 2. Place the dental impression holder on the 3D bite block support 3. Place the black foam on the dental impression holder 4. Click 62 Chapter 8 Acquiring 3D Object Images...
  • Page 69: Launching The Appliance Acquisition

    Note: You will be able to see the object in live-view mode on the workstation screen during the acquisition. When the acquisition ends, the acquired image appears on your screen. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02 63...
  • Page 70 64 Chapter 8 Acquiring 3D Object Images...
  • Page 71: Maintenance Actions

    To maintain optimum image quality, the image quality must be controlled once a month. To control the image quality, follow these steps: 1. On your desktop, double-click . The CS 3D Technician Tools window is 8200 displayed. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 72 2. In the left pane, double-click Image Quality Tool to launch the Image Quality Control software. Click Next to access the Image Quality Control window. 4. Click test options to select the desired test or to modify default test parameters. 66 Chapter 9 Maintenance...
  • Page 73 Click Start a new test, and follow the on-screen instructions. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 74 68 Chapter 9 Maintenance...
  • Page 75: Quick Troubleshooting

    Start the acquisition The acquisition has been Keep pressing and holding the exposure button. launched. Upgrade of the firmware The system is being upgraded. Wait for completion of the upgrade. Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 76 70 Chapter 10 Troubleshooting...
  • Page 77 Hedelfinger Str. 60, 70327 Stuttgart, Germany Carestream Dental Spain S.L.U. Paseo de la Castellana, 79, Madrid 28046, España Carestream Dental Italy S.r.l. Via Mario Idiojmi 3/3, Assago 20090 (MI), Italia Panoramic and 3D Modality User guide for CS 8200 3D (SMA65) Ed02...
  • Page 78 List of Importers for Switzerland CURADEN AG Riedstrasse 12 CH-8953 Dietikon Switzerland Dema dent AG Furtbachstrasse 16 CH-8107 Buchs Switzerland Jordi Röntgentechnik AG Dammstrasse 70 CH-4142 Münchenstein Switzerland E. Schweizer AG Bernerstrasse Nord 182 CH-8064 Zürich Switzerland 72 Chapter 11 Contact Information...
  • Page 79 For more information visit: www.carestreamdental.com 2022-02 SMA65 Ed02...