MR MSENSE DGA 2 Operating Instructions Manual

MR MSENSE DGA 2 Operating Instructions Manual

Online oil analysis
Table of Contents

Advertisement

Online Oil Analysis
®
MSENSE
DGA 2/3
Operating Instructions
BA4001150/10 EN

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MSENSE DGA 2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for MR MSENSE DGA 2

  • Page 1 Online Oil Analysis ® MSENSE DGA 2/3 Operating Instructions BA4001150/10 EN...
  • Page 2 © All rights reserved by Messko GmbH Dissemination and reproduction of this document and use and disclosure of its content are strictly prohibited unless expressly permitted. Infringements will result in liability for compensation. All rights reserved in the event of the granting of patents, utility models or designs.
  • Page 3: Table Of Contents

    Table of contents Table of contents Introduction.........................  6 Manufacturer............................ 6 Subject to change without notice ...................... 6 Completeness.............................  6 Safekeeping............................ 6 Notation conventions ..........................  7 1.5.1 Hazard communication system .......................... 7 1.5.2 Information system.............................. 8 1.5.3 Instruction system .............................. 8 1.5.4 Typographic conventions ............................ 9 Safety..........................
  • Page 4 Table of contents Mounting the device..........................  25 6.2.1 Preparing the transformer and ball valve ...................... 25 6.2.2 Mounting the MSENSE® DGA 2/3........................ 30 Electrical connection.........................  39 6.3.1 Electromagnetic compatibility.......................... 40 6.3.2 Cable recommendation ............................ 40 6.3.3 Routing and preparing the cables ........................ 40 6.3.4 Supply voltage and protective conductor ...................... 41 6.3.5 Device grounding .............................. 42 6.3.6...
  • Page 5 Table of contents 10.3 Oil extraction.............................  70 10.4 Device field calibration........................ 71 10.4.1 Taking a sample for field calibration........................ 71 10.4.2 Field calibration .............................. 73 10.4.3 Reading out the service database........................ 78 10.5 Cleaning............................ 79 Removal.......................... 80 11.1 Removing the MSENSE® DGA 2/3 .................... 80 Disposal..........................
  • Page 6: Introduction

    1 Introduction 1 Introduction This technical document contains detailed descriptions on the safe and proper installation, connection, commissioning and monitoring of the prod- uct. This technical document is intended solely for specially trained and autho- rized personnel. 1.1 Manufacturer The product is manufactured by: Messko GmbH Gewerbegebiet An den Drei Hasen Messko-Platz 1...
  • Page 7: Notation Conventions

    1 Introduction 1.5 Notation conventions This section contains an overview of the symbols and textual emphasis used. 1.5.1 Hazard communication system Warnings in this technical file are displayed as follows. 1.5.1.1 Warning relating to section Warnings relating to sections refer to entire chapters or sections, sub-sec- tions or several paragraphs within this technical document.
  • Page 8: Information System

    1 Introduction Pictograms warn of dangers: Pictogram Definition Warning of a danger point Warning of dangerous electrical voltage Warning of combustible substances Warning of a tipping hazard Warning of a hot surface Table 2: Pictograms used in warning notices 1.5.2 Information system Information is designed to simplify and improve understanding of particular procedures.
  • Page 9: Typographic Conventions

    1 Introduction Aim of action ü Requirements (optional). ► Step 1 of 1. ð Result of step (optional). ð Result of action (optional). Multi-step instructions Instructions which consist of several process steps are structured as follows: Aim of action ü Requirements (optional). 1.
  • Page 10: Safety

    2 Safety 2 Safety This technical document contains detailed descriptions on the safe and proper installation, connection, commissioning and monitoring of the prod- uct. ▪ Read this technical file through carefully to familiarize yourself with the product. ▪ This technical file is a part of the product. ▪...
  • Page 11: Fundamental Safety Instructions

    2 Safety 2.2 Fundamental safety instructions To prevent accidents, disruptions and damage as well as unacceptable ad- verse effects on the environment, those responsible for transport, installa- tion, operation, maintenance and disposal of the product or parts of the prod- uct must ensure the following: Personal protective equipment Loosely worn or unsuitable clothing increases the danger of becoming...
  • Page 12: Personnel Qualification

    2 Safety Ambient conditions To ensure reliable and safe operation, the product must only be operated under the ambient conditions specified in the technical data. ▪ Observe the specified operating conditions and requirements for the in- stallation location. Modifications and conversions Unauthorized or inappropriate changes to the product may lead to personal injury, material damage and operational faults.
  • Page 13: Personal Protective Equipment

    If maintenance is not carried out by our Technical Service department, please ensure that the personnel who carry out the mainte- nance are trained and authorized to do so by Maschinenfabrik Reinhausen GmbH. MR Service & Complaint Maschinenfabrik Reinhausen GmbH Falkensteinstrasse 8 93059 Regensburg Germany service@reinhausen.com...
  • Page 14 2 Safety Personal protective equipment to be worn at all times Protective clothing Close-fitting work clothing with a low tearing strength, with tight sleeves and with no protrud- ing parts. It mainly serves to protect the wearer against being caught by moving machine parts. Safety shoes To protect against falling heavy objects and slip- ping on slippery surfaces.
  • Page 15: Security

    3 IT security 3 IT security Observe the following recommendations for the secure operation of the product. ▪ Ensure that only authorized personnel have access to the device. ▪ Only use the device within an ESP (electronic security perimeter). ▪ Ensure that the device is only operated by trained personnel who are fa- miliar with IT security.
  • Page 16: Product Description

    4 Product description 4 Product description The detection of dissolved gases in transformer oil is generally considered to be the first indication of emerging faults in transformers. Several interna- tional standards highlight the relevance of this examination method, includ- ing the IEEE Std C57.104™-2008, IEC 60422, IEC 60567 and IEC 60599 standards.
  • Page 17: Scope Of Delivery

    4 Product description The device is available in two device versions: Measured variables Hydrogen (H2) Carbon Moisture Oil tempera- monoxide (H2O) ture (CO) MSENSE® DGA 2 MSENSE® DGA 3 MSENSE® DGA 2 for measuring and analyzing the fault gas hydrogen (H2) present in the transformer oil, for detecting the oil moisture content and for measuring the oil temperature.
  • Page 18: Function Description

    4 Product description 4.2 Function description The MSENSE® DGA 2/3 measuring head is positioned in the transformer in- sulating oil. There is often a large distance between the measuring head and the origin of the gases. It is therefore important for the early detection of gases that the oil at the installation position is in motion, either due to natural convection or due to artificially generated oil circulation.
  • Page 19: Design

    4 Product description 4.3 Design Figure 1: MSENSE® DGA 2/3 1 Ventilation 2 Housing cover screw connection 3 Ball valve locking lever 4 Display 5 ENTER key 6 DOWN key 7 UP key 8 Supply voltage cable screw con- nection 9 Signaling relay cable screw con- 10 Analog outputs cable screw con- nection nection...
  • Page 20: Safety Markings And Nameplate

    4 Product description 4.4 Safety markings and nameplate The following safety markings are used on the product: Figure 2: Safety markings and nameplate Description Warning sign "Do not close!" on the slide valve Nameplate Observe the documentation Observe the documentation: See the note about the oil extraction adapter in the chapter "Installing MSENSE®...
  • Page 21: Packaging, Transport And Storage

    5 Packaging, transport and storage 5 Packaging, transport and storage 5.1 Purpose The packaging is designed to protect the packaged product during transport, loading, unloading and during periods of storage in such a way that no detri- mental changes occur. The packaging must protect the goods against per- mitted transport stresses such as vibration, knocks and moisture (rain, snow, condensation).
  • Page 22 5 Packaging, transport and storage Every delivered shipment must be checked for the following by the recipient before acceptance (acknowledgment of receipt): ▪ Completeness based on the delivery slip ▪ External damage of any kind. The checks must take place after unloading, when the box or transport con- tainer can be accessed from all sides.
  • Page 23: Storage Of Shipments

    5 Packaging, transport and storage 5.5 Storage of shipments When selecting and setting up the storage location, ensure the following: ▪ Store the product and accessories in the original packaging until installa- tion. ▪ Protect stored goods against moisture (rain, flooding, water from melting snow and ice), dirt, pests such as rats, mice, termites etc.
  • Page 24: Installation

    6 Installation 6 Installation This chapter describes how to install and connect the device correctly. DANGER Electric shock! Risk of fatal injury due to electrical voltage. Always observe the following safety regulations when working in or on electrical equipment. ► Disconnect the system. ►...
  • Page 25: Mounting The Device

    6 Installation Less recommended installation positions On the side in the lower area; so long as oil circulation is ensured at the in- tended position, installing the device here may be considered. Connection pipe from the transformer tank to the cooling system: the oil tem- peratures here can be very high and influence the quality of the measure- ments Impermissible installation positions...
  • Page 26 6 Installation 2. Remove the fixing screw from the dummy plug. Figure 5: Unchaining the dummy plug 3. Release the clamp connection. Figure 6: Releasing the clamp connection ® MSENSE DGA 2/3 BA4001150/10 EN Messko GmbH 2020...
  • Page 27 6 Installation 4. Pull the dummy plug out. Figure 7: Pulling the dummy plug out Package the dummy plug and store safely in case the MSENSE® DGA is to be removed at a later date. 5. Turn the ball valve lever to the "closed" position. Figure 8: Closing the ball valve Missing or incorrectly positioned round gasket NOTICE...
  • Page 28 6 Installation 6. Lubricate the two round gaskets at the front of the ball valve. Figure 9: Lubricating the ball valve round gaskets 7. Lubricate the two round gaskets at the rear of the ball valve. Figure 10: Lubricating the ball valve round gaskets NOTICE!  Mounting errors can lead to non-tight seals and thus uncon- trolled oil leakage.
  • Page 29 6 Installation 9. Mount the closed ball valve on the transformer and align above the elon- gated holes such that the locking lever is at the top. Use fastening materi- als made of stainless steel A4 in the strength category -70. Figure 11: Mounting the closed ball valve Flange Bolt...
  • Page 30: Mounting The Msense® Dga 2/3

    6 Installation 6.2.2 Mounting the MSENSE® DGA 2/3 1. Remove the device from the packaging and set it down securely on a scratch-resistant and non-slip level surface. Figure 12: Setting the device down securely When doing so, ensure that the display of the device version with display is not damaged.
  • Page 31 6 Installation Store the protective cap in case the device is to be removed at a later date. 3. Open the connection for taking oil samples. To do so, push the bayonet connector towards the device, hold down and pull the sealing plug out. Release the bayonet connector.
  • Page 32 6 Installation NOTICE!  Make sure the oil extraction adapter belongs to this device. If the standard version and offshore version are mixed up, the oil extraction connection and adapter can get damaged. Push the supplied oil extraction adapter into the opening until it audibly snaps into place. Figure 15: Connecting the oil extraction adapter 5.
  • Page 33 6 Installation NOTICE Risk of damage The device measuring pipe can become tilted when it is inserted into the ball valve. ► Support the weight of the device by hand and push the pipe into the ball valve with even pressure up to the stop (tangible resistance). Figure 17: Inserting the device into the ball valve 7.
  • Page 34 6 Installation Risk of burns CAUTION Hot oil can push the device out of the ball valve and spray out. ► Ensure that the safety chain is correctly bolted in place and does not sag. 8. Position a collecting tray below the valve opening in the oil extraction adapter and open the ball valve by aligning the lever up to the stop paral- lel to the measuring pipe.
  • Page 35 6 Installation 11. Close the oil extraction adapter valve. Figure 20: Closing the oil extraction adapter 12. Remove the adapter from the device. To do so, push the bayonet con- nector towards the device, hold down and pull the adapter out. Release the bayonet connector.
  • Page 36 6 Installation 13. Put the sealing plug back on the bayonet connector for venting and tak- ing oil samples and push it in until it snaps into place. Figure 22: Inserting the sealing plug Ensure that the connection is tightly sealed. 14.
  • Page 37 6 Installation Electric shock WARNING Risk of fatal injury due to electrical voltage. ► Maintain the necessary distance from the active part of the transformer (see Device dimensions with 285 mm measuring pipe length [►Section 14.1, Page 93] and Device dimensions with 507 mm measuring pipe length [►Section 14.2, Page 94]).
  • Page 38 6 Installation 16. Reduce the length of the sagging safety chain by inserting the supplied chain connector through two links in the chain and closing it. Figure 25: Locking the safety chain Alternatively, a padlock can be used to secure the device against unautho- rized removal.
  • Page 39: Electrical Connection

    6 Installation 6.3 Electrical connection This chapter describes the correct electrical connection of the device. WARNING Electric shock Risk of fatal injury due to electrical voltage. ► De-energize the device and system peripherals and take measures to en- sure that they cannot be switched back on. Only connect the device to circuits with an external overcurrent protection device and an all-pole isolating device so that the equipment can be fully de- energized if required (for service, maintenance, etc.).
  • Page 40: Electromagnetic Compatibility

    6 Installation 6.3.1 Electromagnetic compatibility The device has been developed in accordance with applicable EMC stan- dards. The following points must be observed in order to maintain the EMC standards. ▪ The system's overvoltage protection must be effective. ▪ The system's ground connection must comply with all technical regula- tions.
  • Page 41: Supply Voltage And Protective Conductor

    6 Installation To prepare the cable correctly, proceed as follows: 1. Open the device connection area. To do so, unscrew the 4 captive screws on the housing cover. The cover is connected to the device via hinges and can be flipped open. 2.
  • Page 42: Device Grounding

    6 Installation Ensure that the protective conductor cross-section is at least the same as all other connection conductors. 6.3.5 Device grounding Connect the ground connection on the device ball valve to the transformer ground connection. Figure 29: Ground connection With offshore devices, the grounding point of the ball valve is to be greased (see chapter Ensuring offshore capability [►Section 6.4, Page 48]).
  • Page 43 6 Installation The terminal strip is assigned by default (factory setting) as follows: Figure 30: Analog outputs ▪ CO concentration: terminals 4, 5, 6 (MSENSE® DGA 3 only) ▪ H2 concentration: terminals 7, 8, 9 ▪ RH relative moisture: terminals 10, 11, 12 To connect the cables, proceed as follows: 1.
  • Page 44: Main Switching Contacts

    6 Installation 6.3.7 Main switching contacts The device has 5 floating main switching contacts: S1, S2, S3, S4 and one safety switching contact = fail-safe relay (FSR) for status signaling. The con- tactors are designed as changeovers and are connected to the terminals 13/14/15 (S1), 16/17/18 (S2), 19/20/21 (S3), 22/23/24 (S4) and 25/26/27 (FSR).
  • Page 45: Scada Connection

    6 Installation To connect the system periphery cables to the main switching contacts, pro- ceed as follows: ü Only use specified cables. Note the cable recommendation. 1. Connect the leads that are to be wired to the device in accordance with the illustration of the main switching contacts.
  • Page 46 6 Installation Device address: 1 to 247 Baud rate: 4800, 9600, ...115200 Parity: even Detailed information is available in the MESSKO® MSET parameterization software operating instructions. 6.3.8.3 Modbus RTU protocol The data point table for the Modbus RTU protocol is available in the Appen- dix [►Section 14.5, Page 98].
  • Page 47 6 Installation 6.3.8.5 Integration of the device into a 2-conductor system Figure 34: 2-conductor system *) When integrating the device into a 2-conductor system, the following con- ductors are to be bridged in the external supply line: ▪ Conductors TXD0 (pin 1) and RXD0 (pin 4) ▪...
  • Page 48: Ensuring Offshore Capability

    6 Installation 6.4 Ensuring offshore capability To prevent corrosion in offshore applications, greasing the following points is recommended (tube of grease included in delivery): 1. Cover the housing cover screws with grease. Time period: After opening the housing cover; at least every 2 years. Figure 35: Greasing the housing cover screws ®...
  • Page 49 6 Installation 2. Fill the holes in the housing base with grease. Time period: At least every 2 years. Figure 36: Holes in the housing base 3. Apply grease to the ball valve steel clamping ring liberally. Avoid greasing the interior of the ball valve. Time period: After each mounting or removal process;...
  • Page 50 6 Installation Figure 38: Grounding point on the ball valve ® MSENSE DGA 2/3 BA4001150/10 EN Messko GmbH 2020...
  • Page 51: Commissioning

    7 Commissioning 7 Commissioning NOTICE Damage to the device! Damage to the electronics of the device due to incorrect supply voltage! ► Provide the correct supply voltage in accordance with the nameplate on an external isolating device. The device is ready for operation once it has been connected to the supply voltage and completed a run-in period of at least 24 hours.
  • Page 52: Installing The Messko® Mset Parameterization Software

    7 Commissioning 2. Connect the device service adapter USB plug to the PC or laptop on which the MESSKO® MSET parameterization software is installed (see corresponding operating instructions). Figure 39: USB service adapter The MSENSE® DGA 2/3 can now be parameterized via the MESSKO® MSET parameterization software.
  • Page 53: Commissioning In Existing Systems

    7 Commissioning 7.3 Commissioning in existing systems Messko GmbH recommends a field calibration of the device parameters to ensure optimum operation, in particular with: ▪ Aged insulating oils ▪ Modified oils (e.g. the addition of additives) ▪ Oils that do not satisfy the standards ASTM D3486-091, IEC 60296 or IEC 60422.
  • Page 54: Settings For Carbon Monoxide (Dga 3 Only), Hydrogen And H2O Concentrations In The Oil

    7 Commissioning ▪ Modbus address ▪ Service database ▪ Field calibration ▪ Default calibration 7.4.1 Settings for carbon monoxide (DGA 3 only), hydrogen and H2O concentrations in the oil The upper warning and alarm values for the carbon monoxide content (CO) (MSENSE® DGA 3 only), the hydrogen content (H2) and the H2O concen- tration can be specified.
  • Page 55: General Settings

    7 Commissioning *) Recommended values based on IEEE C57.104, Condition 1 Humidity (H2O) limit values Limit value Default Minimum Maximum Concentration too high warning 30% RH 0% RH 100% RH Concentration too high alarm 45% RH 0% RH 100% RH Gas formation rate too high warning 10% RH/d 0% RH/d 100% RH/d Gas formation rate too high alarm...
  • Page 56: Modbus Settings

    7 Commissioning Service code Proceed in accordance with the description in the MESSKO® MSET param- eterization software operating instructions in order to enter and transmit a service code. Service Description code No action. Sample extraction: This service code informs the device that an oil sample has been taken for calibrating the absolute humidity H2O.
  • Page 57 7 Commissioning Modbus baud rate setting The transmission speed (baud rate) of the Modbus interface can be set as follows: 4800, 9600, 19200, 38400, 57600, 115200 Bd. Modbus baud rate Standard value 19200 Bd Maximum value 115200 BD Minimum value 4800 BD Take into consideration a minimum pause of 500 ms between two Modbus requests.
  • Page 58: Operation

    8 Operation 8 Operation Observe the difference between the devices versions MSENSE® DGA 2 und MSENSE® DGA 3. Unless stated otherwise, the description refers to the device version MSENSE® DGA 3. Device ver- Measured variables sion Hydrogen (H2) Carbon Moisture Oil tempera- monoxide (H2O) ture (CO) MSENSE®...
  • Page 59: Operating The Device With Display

    8 Operation 8.1 Operating the device with display The display and 3 operating keys can be used to call up and display the set- tings of the device. Use the MESSKO® MSET software included in the de- vice's scope of delivery for the parameterization. There are 3 menu levels: Menu level 1 Menu level 2...
  • Page 60: General Operation

    8 Operation 8.1.1 General operation 8.1.1.1 Menu level 1 = Display of the CO, H2 and moisture content and the oil temperature Action Reaction Press the ENTER key briefly Move forward one menu level 8.1.1.2 Menu level 2 = Setup selection Action Reaction Press UP key...
  • Page 61: Main Screen / Operating Display

    8 Operation 8.1.2 Main screen / Operating display The following displays appear in rotation once the device has been commis- sioned: Figure 41: Display with current H2 and CO content Left side Right side CO (MSENSE® DGA 3 only) Hydrogen content in ppm Carbon monoxide content in ppm Figure 42: Display with current H2O content and oil temperature Left side...
  • Page 62 8 Operation The water content (humidity) display can be converted from %RH to ppm via the MESSKO® MSET parameterization software if required (service code 4 and 5). When the measured data is transmitted (e.g. to a SCADA control center or a laptop), the water content is always transmitted in both display formats.
  • Page 63: Events

    8 Operation These limits and rates are suitable for initial commissioning. However, the systems (transformer and device) must be aligned with each other over time. Thus, with sufficient data, the limits / formation rates for the "warning" and "alarm" values should be able to be calculated and loaded via the MESSKO® MSET parameterization software.
  • Page 64: Operating The Device Without Display

    8 Operation The following table lists the possible events and the respective recom- mended actions: Event Text display Recommended action CO concentration warning CO Warn. (DGA 3 only) CO concentration alarm CO Alarm (DGA 3 only) Gas formation rate CO warn- CO Formation Warn.
  • Page 65: Fault Elimination

    9 Fault elimination This chapter describes how to eliminate operating faults. If a solution cannot be found for a fault, contact Maschinenfabrik Reinhausen GmbH: Maschinenfabrik Reinhausen GmbH MR Service & Complaint Falkensteinstrasse 8 93059 Regensburg Germany E-mail: service@reinhausen.com E-mail: complaint@reinhausen.com Please have the following data to hand: ▪...
  • Page 66: Other Faults

    Message via FSR Cable break in the supply Check wiring, in the event line of doubt contact MR Ser- vice & Complaint Drop in the supply voltage Check supply voltage, in the event of doubt contact MR Service & Complaint...
  • Page 67 9 Fault elimination 3. Remove the safety cover. 4. Using a screwdriver, carefully move under one end of the fuse and lift it out of the holder. 5. Remove the fuse. 6. Position the replacement fuse with both ends on the holder and carefully press down until it locks into place.
  • Page 68: Maintenance

    10 Maintenance 10 Maintenance Inspection and maintenance are required for the operational safety and to maintain measurement accuracy. 10.1 Inspection Monitoring of the device is limited to occasional visual inspections and a reg- ular oil extraction. For efficiency reasons, these inspection tasks can be combined with the usual checks on the transformer.
  • Page 69: Maintenance

    Service Maintenance required Maintenance required Contact MR Technical Service In case of a maintenance signal, please contact the Technical Service of Maschinenfabrik Reinhausen GmbH (MR) immediately. Maschinenfabrik Reinhausen GmbH Technical Service Postfach 12 03 60 93025 Regensburg Germany Phone: +49 941 4090-0 E-mail: service@reinhausen.com...
  • Page 70: Oil Extraction

    10 Maintenance 10.3 Oil extraction Figure 45: Oil extraction The scope of delivery includes two MESSKO® sample extraction kits for tak- ing oil samples via the oil sample extraction connection. For information on taking oil samples, refer to the included instructions BA2938054 and chapter Device field calibration [►Section 10.4, Page 71] in these instructions.
  • Page 71: Device Field Calibration

    10 Maintenance 10.4 Device field calibration The field calibration of the device gas analysis values serves to ensure the correct measurement results: ▪ for aged insulating oils ▪ for modified oils (e.g. the addition of additives) ▪ for oils that do not satisfy the standards ASTM D3486-091, IEC 60296 or IEC 60422.
  • Page 72 10 Maintenance – CO (MSENSE® DGA 3 only) – H2O (%RH or ppm) – UNIX time stamp Figure 46: Complete the sample information sheet (SAMPLE DATA SHEET – OIL ANALYSIS) 2. Take an oil sample in accordance with IEC 60567 and send it to the MESSKO®...
  • Page 73: Field Calibration

    10 Maintenance 10.4.2 Field calibration Once the sample has been analyzed in the oil laboratory, you will be sent the oil sample report for the oil analysis. The following values are important: ▪ Hydrogen H2 ▪ Carbon monoxide CO (MSENSE® DGA 3 only) ▪...
  • Page 74 10 Maintenance The individual analysis values are to be found on the following pages of the analysis report: Figure 49: Analytical values (example) To perform the field calibration, proceed as follows: 1. Connect the MSENSE® DGA 2/3 to the PC via the USB service adapter and open the MESSKO® MSET parameterization software.
  • Page 75 10 Maintenance 3. Click on Service and then on MSENSE field calibration. Figure 51: Service 4. Select the gases for which you want a field calibration. Enter the labora- tory results as the reference corresponding to the respective gas. Figure 52: Field calibration step 1 ®...
  • Page 76 10 Maintenance 5. Enter the date of the oil sample from the laboratory results. Figure 53: Entering the date 6. Click on Start to initiate the field calibration. Figure 54: Starting the field calibration ð The progression of the process can be viewed in the status window. The process may take several minutes.
  • Page 77 10 Maintenance 7. Wait until the status column displays whether or not the calibration is pos- sible. Figure 55: Calibration is possible 8. If the calibration is possible, enter the password and initiate the transmis- sion of the calibration data to the device via Send calibration. Figure 56: Password entry and sending the calibration ®...
  • Page 78: Reading Out The Service Database

    Figure 57: Calibration successful 10. Upon successful calibration, you can close the window and disconnect the connection between the MSENSE DGA 2/3 and the laptop. 11. Unscrew the USB service adapter. 12. Then screw the protective cap onto the M12 socket to protect the device against water ingress.
  • Page 79: Cleaning

    2. Click on Service in the top right. 3. Click on the Create service database button and save the zip file to the laptop. Figure 58: Service database 4. Contact MR Technical Service and send the zip file. Maschinenfabrik Reinhausen GmbH Technical Service Postfach 12 03 60...
  • Page 80: Removal

    11 Removal 11 Removal CAUTION Hot oil escape Risk of injury and slipping. ► Wear safety clothing. ► Carry out the operating steps in the specified sequence. ► Ensure that the gate valve remains closed after the removal of the de- vice.
  • Page 81 11 Removal 2. Release the ball valve clamp connection. Figure 60: Releasing the ball valve clamp connection 3. Pull the device as far as possible out of the ball valve. The chain should now be tightly tensioned. ® Messko GmbH 2020 BA4001150/10 EN MSENSE DGA 2/3...
  • Page 82 11 Removal NOTICE Danger of damaging the measuring pipe! If the device is operated in front of a slide valve, the tightly tensioned chain ensures that the measuring pipe of the device has been pulled out far enough and that the slide valve does not collide with it when closing. Figure 61: Pulling the device out 4.
  • Page 83 11 Removal Danger of crushing due to falling device CAUTION The device is now free. Protect yourself against spraying oil with personal protective equipment and support the weight of the device. 6. Unscrew the fixing screw for the safety chain on the device far enough to free the chain.
  • Page 84 11 Removal NOTICE Type and source of danger Ensure that the device does not become jammed during this procedure. Support the weight by hand. Collect the excess oil in the collecting tray. Figure 64: Pulling the device out of the ball valve ®...
  • Page 85 11 Removal 8. Set the device down securely on a scratch-resistant and non-slip level surface. Figure 65: Setting the device down securely When doing so, ensure that the display of the device version with display is not damaged. 9. Place the supplied red protective cap onto the measuring pipe and close the protective cap opening with adhesive tape.
  • Page 86 11 Removal 10. Clean the device with a dry cloth. The device has now been removed and can be transported or stored in suit- able packaging. The ball valve is then to be secured: Hot oil escape CAUTION Risk of injury and slipping. ►...
  • Page 87 11 Removal 2. Close the air-vent valve and tighten the ball valve clamp connection. Figure 68: Tightening the ball valve clamp connection 3. Secure the free end of the safety change with the dummy plug fixing screw. Attach the chain under as much tension as possible. Figure 69: Securing the dummy plug with link chain ®...
  • Page 88 11 Removal 4. Remove the warning sign "Do not close!" on the slide valve. Figure 70: Slide valve warning sign ® MSENSE DGA 2/3 BA4001150/10 EN Messko GmbH 2020...
  • Page 89: Disposal

    12 Disposal 12 Disposal Observe the national disposal regulations in the respective country of use. ® Messko GmbH 2020 BA4001150/10 EN MSENSE DGA 2/3...
  • Page 90: Technical Data

    DGA measuring range with respect to the oil temperature +10 to +90°C Ambient humidity 0...100%RH Ambient temperature storage/transport -40 to +80°C Please contact MR to check the use in case of extreme conditions ® MSENSE DGA 2/3 BA4001150/10 EN Messko GmbH 2020...
  • Page 91 13 Technical data At the point of measurement Device properties Mechanical connection Via ball valve and configurable flange DN50 or DN80 (several flanges available); with safety chain for safe installation and removal Measuring pipe length Two different lengths for optimum measuring conditions; Short measuring pipe: 285 mm;...
  • Page 92 Setting the time stamp of the oil sample extraction; Entry of the reference values in accordance with the laboratory report Service Reading out the internal device service database for further analysis by MR ex- perts Please note that Microsoft has stopped providing regular support for ver- sions earlier than Windows 10.
  • Page 93: Appendix

    14 Appendix 14 Appendix 14.1 Device dimensions with 285 mm measuring pipe length Figure 71: Device dimensions with 285 mm measuring pipe Flange, see table [►Section 14.3, Page 95] Minimum immersion depth Recommended immersion depth ® Messko GmbH 2020 BA4001150/10 EN MSENSE DGA 2/3...
  • Page 94: Device Dimensions With 507 Mm Measuring Pipe Length

    14 Appendix 14.2 Device dimensions with 507 mm measuring pipe length Figure 72: Device dimensions with 507 mm measuring pipe Flange, see table [►Section 14.3, Page 95] Minimum immersion depth Recommended immersion depth ® MSENSE DGA 2/3 BA4001150/10 EN Messko GmbH 2020...
  • Page 95: Connecting Flange Dimensions

    14 Appendix 14.3 Connecting flange dimensions Figure 73: Flange DN50 PN6; DN50 PN16 For o-ring gasket (included in delivery) For any gasket (not included in delivery) Figure 74: Flange DN80 PN16 For o-ring gasket (included in delivery) For any gasket (not included in delivery) ®...
  • Page 96 14 Appendix Flange dimensions table Flange version mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] DN50 PN6 Ø140 [5.51"] Ø110 [4.33"] 70° 4 x 50° Ø13 [0.51"] Ø81 [3.19"] Ø65 [2.56"] DN50 PN16 Ø165 [6.50"] Ø125 [4.92"] 70° 4 x 50°...
  • Page 97: Electrical Connection

    14 Appendix 14.4 Electrical connection Figure 76: Electrical connection Supply voltage (overvoltage category III) 95...280 V AC 50/60 Hz or 95...280 V DC (any polarity) Protective conductor 4-12 Analog outputs 4…20mA passive (must be supplied with 24 V DC) 13-27 Main switching contacts (crossovers for state signaling): capacity: 30 V DC / 5 A to 300 V DC / 0.25 A or 250 V AC / 5 A;...
  • Page 98: Data Point Table For Modbus Rtu

    14 Appendix 14.5 Data point table for Modbus RTU Separate input register Address Size Description Designation 1 byte On / Off CO limit 1 1 byte On / Off CO limit 2 1 byte On / Off H2 limit 1 1 byte On / Off H2 limit 2...
  • Page 99 14 Appendix Address Size Description Designation (dec.) 4 bytes, Empty 2 SFLOAT 4 bytes, None H2O %RH SFLOAT 4 bytes, Empty 4 SFLOAT 4 bytes, None H2O rate %RH/d SFLOAT 4 bytes, None CO rate ppm/d SFLOAT 4 bytes, None H2 rate ppm/d SFLOAT 4 bytes,...
  • Page 100: Spare Parts List

    14 Appendix 14.6 Spare parts list Seq. Art. No. Designation 1001099900 MSENSE® 2/3 oil extraction unit (standard) Sample extraction adapter (standard version) Material: stainless steel V2A and PTFE 1001100100 MSENSE® 2/3 oil extraction unit (offshore) Sample extraction adapter (offshore version) Material: stainless steel V4A and PTFE 1001100300 MSENSE®...
  • Page 101 14 Appendix Seq. Art. No. Designation 1001101000 MSENSE® 2/3 ball valve DN80 PN16/285 For measuring pipe length 285 mm Flange diameter: DN80 Pressure rating: PN16 Material: stainless steel V4A Including O-ring gasket, material: FPM (Viton) 1001101100 MSENSE® 2/3 ball valve DN80 PN16/507 For measuring pipe length 507 mm Flange diameter: DN80 Pressure rating: PN16...
  • Page 104 Messko GmbH Gewerbegebiet An den Drei Hasen Messko-Platz 1 61440 Oberursel, Germany +49 (0)6171 6398 0 messko-info@reinhausen.com www.reinhausen.com/messko ® BA4001150/10 EN - MSENSE DGA 2/3 - 1000590110 - 06/20 - Messko GmbH 2020 THE POWER BEHIND POWER.

This manual is also suitable for:

Msense dga 3

Table of Contents