Furnace Backwards Flipping - Katanax K2 Prime Manual

Automatic fluxer
Table of Contents

Advertisement

1. Disconnect the instrument from the wall outlet.
2. Lift the safety shield securely up/backwards.
3. Ask for the assistance of a helper.
4. While the helper is maintaining the door lever, remove its three (3) front
5. With the helper now maintaining both levers, remove other side's screws.
6. Have your helper slowly accompany the levers' upwards movement until the
Furnace
backwards
flipping
To access components located in the base, it is necessary to flip the furnace to the
back:
1. Disconnect the instrument from the wall outlet.
2. Lay the instrument on a surface large enough to accommodate the back face
3. Manually open the door, and pull the platinumware holders out of the furnace.
82
WARNING: Heating elements are powered at high voltage. Always disconnect
the instrument from the wall outlet before removing door. When no door is
present, safety interlocks are bypassed.
screws.
levers stop.
of the furnace when you will flip it back.