This section describes tasks that are performed on a regular basis, and are
performed to adjust or repair a malfunctioning system of the instrument.
Should you have any question, or need further assistance, please do not hesitate
to contact us (see page 91, Contacting Katanax).

Warning

Platinumware
holders service
This procedure describes the steps to take in order to remove your K2 Prime
platinumware holders. This procedure needs to be done when cleaning, or when
replacing the holders or parts of it.
Note that both holders (crucible and mold) can be removed totally or partially. In
routine cleaning, only the middle portions would need to be removed; you would
not need any tool. However, if more parts are damaged or soiled, then the whole
holders must be removed, and tools are needed. Please refer to page 68,
Adjustment of holders, to properly re-adjust the holders after maintenance.
Also, because the crucible holder sits above the mold holder, we recommend
removing the crucible holder first, and then the mold holder.
Crucible
holder removal
1. Crucible holder removal – middle section only
60

Service operations

IMPORTANT: Some of the procedures described in the following pages imply a
risk of death by electrocution; those procedures shall be executed only by
trained personnel.
1.1
Remove any crucibles and molds from the holders.
1.2
Remove the front ceramic bar of the mold holder.
IMPORTANT: The middle section of the crucible holder must remain gently
compressed longitudinally at all times, or the ceramic rods, tubes and spacers
will fall. Using the elastic band with two clips, originally provided with the
complete crucible holder, is recommended for this operation.