Making Solutions - Katanax K2 Prime Manual

Automatic fluxer
Table of Contents

Advertisement

7. Add the non-wetting agent if it is not already integrated within the flux. Solid
8. Place crucibles in the holders. Important: make sure that they are properly
9. Place molds on their holder. Important: do not forget to install a mold under
10. Touch Start to launch the fusion. If the preset temperature is not yet reached
11. Upon heating, the sample reacts if an oxidizer is present. Then, the flux melts
12. At the end of the cycle (when the cooling blower turns off), carefully pick up
Making
solutions
When making a solution, the process is quite similar, but the mold installation
changes for the following:
9. Place beakers in instrument.
The magnetic agitation system is always active.
30
6.2
Mix thoroughly with the sample. (Try to leave the bottom layer of flux
untouched.)
non-wetting agent should be thoroughly mixed with the flux. Making an
aqueous solution with the solid salts can also prove very convenient, and can
be pipetted on top of the dry ingredients.
installed. See page 27, Crucible installation, for details.
each crucible. See page 29, Mold installation, for details.
in the furnace, a few minutes' delay will allow sufficient heating, then the
holders will automatically enter the furnace. The door opens and closes
automatically.
and dissolves the sample. The mold is heated at the same time. At the end,
the door opens to let the crucibles out, and those are tilted to pour into the
molds. A blower cools the disks.
the glass disks. Important: They may still be very hot, depending on the
mold weight and program parameters.
IMPORTANT: Before attempting to make solutions, it is important to remove
the middle section of the mold holder, which can otherwise interfere with the top
of the beakers.
IMPORTANT: Katanax does not recommend attempting to prepare solutions in
a fluxer not fitted with the optional solution magnetic stirrers.
9.1
Fill PTFE beakers with about 100 ml suitable dilute acid. (Nitric acid is the
most commonly used.)
9.2
Add magnetic stirring bars.
9.3
Press the "load beaker" button to slide the platinumware
holders back, against the furnace door.
9.4
Put the beakers on the agitator, in their respective holes.
9.5
Optionally, press the "load beaker" button again, to
move the platinumware holders to the front and clear
the door opening.

Advertisement

Table of Contents
loading

Table of Contents