The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
3 V 0.15 MHz to 80 MHz
Conducted RF
6 V in ISM and amateur
IEC 61000-4-6
radio bands between
0.15 MHz and 80 MHz
Radiated RF
10V/m 80 MHz to 2.7
IEC 61000-4-3
GHz
Table 3
Declaration - Electromagnetic Immunity
Compliance Level
3 V 0.15 MHz to 80
MHz 6 V in ISM and
amateur radio bands
between 0.15 MHz and
80 MHz
10V/m
Electromagnetic Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the device than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter. Recommended
separation distance
d = 1.2 √P 150 KHz to 80 MHz
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 80 MHz to 2.7 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.