Final activities

8Remaining work

Protective conductor test

Observe the protective conductor resistance test information in these instructions.
Perform the protective conductor test with all covers closed in accordance with
ARTD-002.731.17... The protective conductor resistance must not exceed 0.2 Ohms.
P
The values determined must be recorded as initial measurements in the protective con-
ductor resistance report (chapter 11 of these instructions), and the measuring points
must be given.
P
The measuring procedure and the measuring device used (designation and serial num-
ber) must also be documented.

Leakage current measurement

Observe the leakage current measurement information in these instructions.
The leakage current must be measured with all covers closed, in accordance with
ARTD-002.731.17... The limit value of 2.5 mA must not be exceeded.
WARNING
The system must be switched on during measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
P
Enter the highest value as the first measured value in the leakage current report (chap-
ter 11 of these instructions).
This value must not exceed the permissible leakage current values according to DIN
VDE 0751-1/2001-10, Table F.1, line "leakage current for devices according to re-
marks 1 and 3", of 2.5 mA.
P
Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, correct the measured value to the value corresponding to a mea-
surement at the nominal value of the line voltage. This must also be documented.
P
The measuring procedure (differential measurement or direct measurement) and the
measuring instrument used (designation and serial number) must also be documented.

Patient leakage current measurement

Observe the patient leakage current measurement information in these instructions.
The patient leakage current must be measured with all covers closed, in accordance
with ARTD-002.731.17...
MOBILETT XP Digital
Electrical voltage!
Non-compliance can lead to severe injury and even death.
No housing parts of the system may be touched during
direct measurement of the leakage current .
Third-person access to the system must be prevented.
SPR8-230.814.30.03.02
11.05