Protective Conductor Test; Device Leakage Current Measurement - Siemens POLYMOBIL Plus Start-Up Instructions

Hide thumbs Also See for POLYMOBIL Plus:
Table of Contents

Advertisement

Final procedures
Fig. 1

Protective conductor test

Observe the information on the protective conductor test in these instructions.
Perform the protective conductor test with the system closed according to ARTD-
002.731.17...
The protective conductor resistance must not exceed 0.2 ohms.
Record the determined values as first measured values in the report 'Protective
conductor resistance / report' (Chapter 6 of these instructions) stating the measuring
points.
In addition document the measuring method and the measuring instrument used
(designation and serial number).

Device leakage current measurement

Observe the information on the device leakage current measurement in these instruc-
tions.
Perform the device leakage current measurement with the system closed according to
ARTD-002.731.17...
The limit of 2.5 mA must not be exceeded.
WARNING
Siemens AG
Medical Solutions
Electrical voltage!
Non-compliance can lead to severe injuries up to death.
In the direct measurement of the device leakage current no hous-
ing parts of the system may be touched during the measurement.
Access to the system for third persons must be prevented.
SPR8-125.034.02
Rev. 02
Page 1 of 2
03.05
CS PS 24
4 - 1
4
4
4