Requirements for Installation

WARNING!
1. Always ground this unit
 Be sure to connect the earth wire (the green cable of power cord) to the grounding conductor
or ground terminal to prevent accidents caused by electric leakage.
 Do not connect the earth wire to gas or water pipes. If not, fire disaster may be caused.
 Do not connect the earth wire to the ground for telephone wire or lightning conductor. If not,
fire disaster or electric shock may be caused.
 Do not use a branching receptacle, which may cause the heat generation.
2. Choose a proper place for installation
 Do not install this unit in a place where:
 Rough or dirty surface.
 Flammable gas or corrosive gas is generated.
 Ambient temperature exceeds 35C.
 Ambient temperature fluctuates violently.
 There is direct sunlight.
 There is excessive humidity and dust.
 There is a constant vibration.
 Install this unit on a stable place with the space as shown below.
3. Caution at defrost
 Catch drain from drain hose with a tray at defrost.
Note: The tray is not included in the attachments of unit.
More than 15cm
More than
15cm
4

Before Using this unit

More than
15cm
More than 1m
Front side
Drain hose
ドレインホース
水受容器
Tray