Input Program - Yamato IN604 Instruction Manual

Programmable low temperature incubator
Table of Contents

Advertisement

Input Program

Select program function
1
 Check that the power in turned on.
Select step
Program OPR
Edit step
2
 The following four setting items are included in a step.
Setting item
Setting temperature
Setting time
Repeat initiating step
Repeat count
 The unit operates with fixed rate inclination if the "START TEMP" and the "Set Temp" are different.
It enters into "Wait" if the temperature does not reach to the setting temperature within a setting
time.
PROGRAM
S01
TROUBLE
ENTER
CANCEL
TEMP.
START.
TEMP.
(Operation with fixed rate inclination)
① Press the PROGRAM key. The program menu
starts and step number selection screen is
displayed.
Press the "▼▲". The registered step numbers
and the smallest number of un-used step are
displayed in sequence. Select the step number
among them.
 The "S01" is displayed when no program is
registered. In this case, the "▼▲" are invalid.
② Press the ENTER key.
number is decided and the setting item selection
screen is displayed.
displayed first.
③ Press the CANCEL key to cancel the program
menu.
Setting range
Setting temperature range by product type
0 minute to 999 hour and 59 minutes, End
None, Registered step numbers 1 to 32
1 to 9999, Infinity
Set Temp.
Set Time
27
Operation Method
The selected step
The "Set TEMP" is
TIME