3.2.

Troubleshooting the Panel PC

3.2.1. Replacing the battery for the RTC (Real Time Clock) on the Panel PC

Tools needed: Phillips Screwdriver, #2
Unplug all the cables and power cords going to the Panel PC. Make note of
what ports cables are connected to.
Remove the Panel PC from the cart and place it with the screen facing down
on a table.
Remove the 11 M3 x 10mm Phillips screws (Natus P/N 800213) that hold the
outer shell on and remove the outer shell.
FIGURE 17: Location of Panel PC outer shell screws
ALGO® 5 Newborn Hearing Screener Service Manual
P/N 000976 Rev.A