2.

Routine Maintenance Procedure

This chapter defines procedures for routine maintenance of the ALGO 5 screener. These
procedures must be performed after any of the following events and/or as prescribed by
the institution's maintenance schedule:
Institution first receives the ALGO 5 screener.
Foreign material or liquid has been deposited on the ALGO 5 screener.
ALGO 5 screener has been visually damaged or subjected to mechanical
shock (e.g. being dropped).
ALGO 5 screener has been submitted for maintenance or scheduled
performance verification.
2.1.

Inspecting Accessory Parts

Routinely inspect screener components as described below:
Inspect earphone cable for damage (tears, cracks, exposed wires, etc.)
Inspect sensor cable for proper connection to the Preamp 5, for worn or
broken insulation, and for looseness at the clips. Inspect Preamp 5 cable for
damage.
Inspect power cord for fraying or other damage.
If the screener is frequently moved to different locations or subjected to heavy
use, check regularly for external damage.
Examine the Preamp 5 for cracks or dents that might have been caused by
repeated banging against the side of the screener or the edge of a crib.
Replace accessories or have them repaired as needed.
Warning! Do not use the screener if you suspect that it is not functioning properly,
or if any parts appear damaged. Contact Natus Technical Service or your
authorized service representative.
2.2.

Cleaning the ALGO 5

2.2.1. Cleaning the Screener

WARNING
To keep the screener clean, perform the following tasks regularly:
Remove dust from the screener's exterior with a soft brush or cloth.
Clean the exterior case of the screener with a lightly water-dampened soft
cloth.
To clean the Panel PC, dilute a volume of glass cleaner with an equal volume
of water. Spray a small amount of the diluted cleaner on a soft cloth and
gently wipe the screen.
12
Turn off power and unplug the screener before cleaning.
ALGO® 5 Newborn Hearing Screener Service Manual
P/N 000976 Rev.A