Appendix A. CE Mark Compliance

A.1 Introduction
For CE Mark compliance, the Model XGM868i flow transmitter must be wired in accordance with the instructions in
this appendix.
CE Mark compliance is required only for units intended for use in EEC countries.
IMPORTANT:
A.2 Wiring
The Model XGM868i must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to Table 13 for the specific requirements.
Connection
Transducer
Armored RG62 a/U
Armored 22 AWG shielded (e.g. Baystate #78-1197)
Input/Output
with armored material added to outside of jacket
Power
Armored 14 AWG 3 conductor
Note: If the Model XGM868i is wired as described in this appendix, the unit will comply with the EMC Directive
2004/108/EC.
DigitalFlow™ XGM868i Startup Guide
Table 13: Wiring Requirements
Cable Type
Appendix A. CE Mark Compliance
Ground Termination
Grounded via cable gland.
Grounded via cable gland.
Grounded via cable gland.
61