The following message appears in the display window.
The calibration sheet as shown below will be printed.
At this moment, the calibration data will be read with the built-in densitometer. Once
calibration has been completed, the message display window goes back to the nor-
mal display ("Ready").
If the calibration pattern cannot be read properly with the built-in
densitometer, the message "Perform Calibration" appears in the
display window. In this case, calibrate the tray again. If the den-
sitometer fails to read the calibration pattern again, contact your
nearest Konica Minolta technical representatives.
51
Calibration