5�1 Maintenance, Items to be Checked, and Schedule
This section describes maintenance and items to be checked necessary to ensure use of DRYPRO
873 in optimum condition.
Maintenance and Items to be Checked by Users
Maintenance and items to be checked by users are as follows: Use the maintenance period as a guide.
Perform maintenance as needed depending on usage environment.
Description of Maintenance Work
Printing QC patterns and
checking QC pattern results
Cleaning film ejection tray
Cleaning supply tray
Cleaning air intakes/outlets
Changing deodorant filter
Cleaning the Cleaning Roller
Maintenance and Items to be Checked by Service Personnel
Other major maintenance and items to be checked are as follows. These are included in the mainte-
nance contract. For performance of maintenance, contact Konica Minolta technical representatives
Maintenance Parts
Supply unit
Pickup roller
Heat process
Opposed rollers (several sets)
unit
Opposed rollers
Cleaning guide
Heat development drive unit
Sub-scanning
Sub-scanning belt
unit
Control box
Air intake filter
Power supply
Air intake filter
unit
*Overseas machines equipped with a transformer have filters and thus need cleaning.
Maintenance Period
When setting a
reference value
In normal times
Replacement
Replacement
Wiping with alcohol
Wiping with alcohol
Replacement
Replacement
Removing dust
Removing dust
102
Weekly
Quarterly
Three times in a row
Every 10,000 prints or every two years
Every 2,000 prints (mammography media users)
Description
Semiannually
Maintenance Period
Every 50,000 prints
Every 50,000 prints
Every 50,000 prints
Every 50,000 prints
Every 200,000 prints
Every 200,000 prints
Every 50,000 prints
Every 50,000 prints