Operating ECal Module
Setting Up a Calibration

Required Procedure for All Calibrations (PNA and PXIe)

For all calibration types, complete the following steps:
1. Connect an ECal module to the analyzer having the appropriate frequency
range and connector type.
2. Wait until LED turns green.
3. Set up the analyzer in the measurement configuration. Select the
frequency, power, sweep and other stimulus settings.
4. View the response (uncorrected) and optimize the analyzer settings as
needed.
5. Connect the ECal module to the measurement ports.
6. Navigate to Response > Cal Wizard
7. In the Calibration Wizard: Begin Calibration window that opens, select Use
Electronic Calibration (ECal), click Next, and follow the prompts.
8. When the calibration is complete, remove the ECal module, and connect
the DUT.
Excessive torque can damage ECal module connectors. See
page 4-21
For optimal results, terminate any unused ECal ports with a 50 ohm load.
Required Procedure for All Calibrations
E5072A)
For all calibration types, complete the following steps:
1. Connect an ECal module to the analyzer having the appropriate frequency
range and connector type.
2. Wait until LED turns green.
3. Set up the analyzer in the measurement configuration. Select the
frequency, power, sweep and other stimulus settings.
4. View the response (uncorrected) and optimize the analyzer settings as
needed.
5. Connect the ECal module to the measurement ports.
6. Select the channel you want to calibrate: Press Channel Next/Channel
Prev keys
7. Press Cal
8. Click ECal > 1-Port Cal
Electronic Calibration Modules Reference Guide
for the required torque setting for each connector type.