Storing and Recalling Data from Memory

The Fisher Scientific accumet AE150 pH benchtop meter allows up to 100 data sets to be
stored and recalled in its non-volatile memory in a Last-In-First-Out (LIFO) sequence. Data
sets are stored with a specified location, as indicated on the display. For example, 32 pH
and temperature data sets and 18 mV and temperature data sets can be stored.
Memory Input:
The memory input feature allows measurement readings to be
stored for later review. This mode can be used at any time during
measurement mode or when the reading is frozen ("HOLD" icon is
shown on the display by pressing the HOLD key).
1. To store a reading during any measurement or HOLD
function, press the MI key to input data into the memory.
2. The "MEM" icon, "StO" and memory number will be shown
briefly on the display.
3. The meter will then return to the measurement mode.
4. To store another reading, press the MI key again.
Note: When the memory is full, the oldest stored data set will be erased to create space for
the new data set.
Memory Recall:
Readings stored in the memory are retained even when the meter is
powered off. To erase all readings stored in memory, use the P1.0
program in the setup mode.
1. In the measurement mode, press the MR key to retrieve the
most recent stored reading.
2. The memory location screen with "MEM" icon, "LOC" and the
memory number will be shown on the display.
3. Press the ENTER key to recall the reading stored under that
memory number.
4. Press the ENTER key again to return to the main memory
recall mode.
5. To view additional data sets, press the  or  arrow key to
scroll to the desired data set location.
a. For example, to view the third data set, press the 
or  arrow key until "LOC 3" is shown and press the
ENTER key to view the data from this location.
6. To continue reviewing additional stored data, press the
ENTER key again.
7. To exit the memory recall mode, press the CAL/MEAS key
to return to the measurement mode.