Instruction Manual
Fisher Scientific
accumet
Basic (AB)
®
Benchtop Meters
AB150 ● AB200 ● AB250
68X613601 Rev 0 July 2012

Advertisement

Table of Contents
loading

Summary of Contents for Fisher Scientific accumet AB200

  • Page 1 Instruction Manual Fisher Scientific accumet Basic (AB) ® Benchtop Meters AB150 ● AB200 ● AB250 68X613601 Rev 0 July 2012...
  • Page 3: Table Of Contents

    Table of Contents 1. Introduction ........................ 3 2. Keypad & Display ....................... 4 Connections: ..........................6 3. System Setup & Configuration ................. 7 STABILITY ..........................7 EXPORT DATA ........................... 7 DATE & TIME ..........................8 PASSWORD ..........................8 FACT RESET ..........................9 CONTRAST ..........................
  • Page 4 11. Conductivity Standardization (Automatic) .............. 17 12. Cond/TDS/Salinity/Resistivity Standardization (Manual Adjustment) ....18 13. Ion Setup ........................20 MEASURE UNIT ......................... 20 ALARM ............................20 STD DUE ............................. 20 STANDARDS ..........................20 14. Ion Standardization ....................21 15. Viewing, Transferring, and Printing Data ..............22 16.
  • Page 5: Introduction

    Introduction Thank you for selecting the Fisher Scientific accumet AB series benchtop meter. All models include an electrode arm with metal bracket (13-637-671), 110/220 universal power supply (13-636-104), RS-232, and USB cable. The electrode arm can be attached to the left, right, or center of the meter according to your preference using a Phillips screwdriver.
  • Page 6: Keypad & Display

    Keypad & Display FIXED KEY DESCRIPTION to initialize and complete standardization (calibration) of displayed parameter MODE to toggle between available measurement modes to enter Setup mode; use to change system settings or SETUP measurement parameters PRINT / ENTER to print, also to confirm selection during setup SAVE / ▲...
  • Page 7 USB port is active. RS-232 port is active. Based on the stability speed selected, instrument has detected a Stable stable reading. Data was sent to the printer; either manually by pressing PRINT key or automatically at timed intervals as entered thru the System Setup mode. “High Alarm”, “Low Alarm”, or “Std Due”...
  • Page 8: Connections

    Connections: Power supply, 110/230 VAC, 9 VDC, 18 W, center positive. Power Use only model 13-636-104 (included). RS-232 serial communication output, 2.5 mm jack. RS-232 For use with RS-232 cable 13-637-681 (included). MINI-B USB communication for field software upgrades (cable included) and data output.
  • Page 9: System Setup & Configuration

    System Setup & Configuration Use the System Setup to customize operation of your AB series meter. Press SETUP from the measurement screen and then press PRINT/ENTER when SYSTEM is highlighted to access these settings. The following settings can be customized for each model: STABILITY STIRRER EXPORT DATA...
  • Page 10: Date & Time

    Timed: When either printing or datalog is ON, the option to select Timed readings ON or OFF appears. Select “ON” to automatically print or save data at selectable intervals. Select “OFF” to save data manually using the PRINT/ENTER key. Time Interval: (MM:SS) (Minutes:Seconds). When timed reading is ON, you must select an interval at which the data will be sent to the Printer and/or Datalog memory.
  • Page 11: Fact Reset

    FACT RESET Select “YES” to reset the to the factory default settings except; Date & Time setting, and data stored into memory. CONTRAST Optimize the contrast setting of your display for best visibility in your surrounding lighting conditions. Test various contrast settings for best results. This setting will be applied to both backlight and non-backlight conditions.
  • Page 12: Setup Ph & Mv

    Setup pH & mV Use Setup pH or mV mode to customize these parameters. Note: mV setup offers Alarm setting only. Press SETUP from the measurement screen and then press ENTER when pH or mV is selected. BUFFER (pH Only) Choose one of (5) pH buffer groups for calibration;...
  • Page 13: Ph Standardization (With Preset Buffer Group)

    pH Standardization (With Preset Buffer Group) For best results, periodic standardization (calibration) with known accurate standards is recommended prior to measurement. Use standards that bracket your intended measuring range while including a neutral point (7.00, 6.86, or 6.79). For example, if you expect to measure samples from pH 6.2 to 9.5, calibration with 4.01, 7.00, and 10.01 will work well.
  • Page 14: Ph Standardization (With Custom Buffers)

    5) Press STD to accept the value of the measured buffer. If successful, the Standardization Report will show the accepted value along with other data. The percentage slope is visible after two or more calibration points have been completed. 6) Repeat Steps 4 & 5 with additional pH buffers. pH Standardization (With Custom Buffers) Buffer Group pH Values...
  • Page 15: Mv Standardization (Offset Adjustment)

    mV Standardization (Offset Adjustment) Oxidization Reduction Potential (ORP or Redox) as measured by an ORP electrode in milliVolt (mV) units is not a precise measurement, but is useful as a relative indicator. As such, mV offset adjustment is not meant to enhance accuracy, but rather to make readings comparable to a reference.
  • Page 16: Temperature Standardization

    Temperature Standardization The thermistor sensor used for automatic temperature compensation and measurement is both accurate and stable, so require frequent calibration isn’t required. Temperature calibration is recommended upon electrode replacement, whenever the temperature reading is suspect, or if matching against a certified thermometer is desired.
  • Page 17: Conductivity, Tds, Salinity, & Resistivity Setup

    Conductivity, TDS, Salinity, & Resistivity Setup Use Setup mode to customize Conductivity, TDS, Salinity, & Resistivity parameters. Press SETUP from the measurement screen and then press ENTER when Conductivity, TDS, Salinity, or Resistivity is selected. STD METHOD (Conductivity Mode Only) Choose automatic or manual standardization for conductivity, and manual standardization for TDS, Salinity, and Resistivity.
  • Page 18: Pure Wat Coef (Pure Water Coefficient)

    PURE WAT COEF (Pure Water Coefficient) Choose ENABLE to automatically apply pure water temperature correction for measurements below 2 µS/cm (ultra pure water). Note: this option does not apply to Salinity mode. TEMP COEFFICIENT (Linear) The temperature coefficient is the amount of change in conductivity per degree temperature (% per ºC).
  • Page 19: Alarm

    ALARM Use a visual and audible alarm to alert you when High or Low values that have been set from this menu are exceeded. “High” or “Low” will blink on the display if the values are exceeded while simultaneously, a loud, intermittent beeping sound is heard.
  • Page 20: Cond/Tds/Salinity/Resistivity Standardization (Manual Adjustment)

    Additional Notes On Automatic Conductivity Calibration: A maximum of one calibration point per range can be performed. If multiple calibration points are used in the same range, the most recent one will replace the previous one. If the electrode is replaced, it is best to clear the calibration to the factory default values.
  • Page 21 4) The upper display is the active reading while the lower display is the factory default value. Use the ▲/▼ keys to adjust the upper display to match the desired value. 5) Press STD after ‘Stable’ appears to confirm the value. The calibrated value, Range (R) that was calibrated and the calculated cell constant are now shown.
  • Page 22: Ion Setup

    Ion Setup Use Setup ISE mode to customize this parameter. Press SETUP from the measurement screen and then press ENTER when ISE is selected. MEASURE UNIT Select ppm, molar, or mg/L ion concentration units. ALARM The AB250 offers a visual and audible alarm to alert you when the High or Low values that have been set from this menu are exceeded.
  • Page 23: Ion Standardization

    Ion Standardization The AB250 can measure ion concentration such as ammonia or fluoride when using an ion selective electrode (ISE) for the specific ion of interest. Ion standardization is required with at least two calibration standards in order for the instrument to calculate the measurement. The primary display will show “- - - -”...
  • Page 24: Viewing, Transferring, And Printing Data

    Viewing, Transferring, and Printing Data To send data directly to a printer or PC you will require the appropriate RS-232 or USB cable.  Press VIEW key to view calibration or stored data directly on the display.  Press REPORT soft key to view calibration data directly on the display. ...
  • Page 25: Standardization Report

    Standardization Report The Standardization Report provides detailed information on the most recent calibration. It includes date & time, calibration standard(s), offset, temperature, number of days calibration is overdue, and slope information. The Standardization Report is automatically displayed every time a standard has been successfully accepted.
  • Page 26: Troubleshooting

    Troubleshooting Parameter Cause Solution Cal points are Custom pH buffer calibration values must too close be at least 1.0 pH unit apart. Reading is 7.00 and Electrode tip may be cracked or broken. won’t change Replace electrode. pH or Ion Slope error / Calibration values are too far apart, Standardization...
  • Page 27: Usb Driver & Boot Loader Documentation

    This document is intended to guide the reader through the process of A) installing the USB driver for communication with a PC, and B) boot loading (upgrading) the ® latest firmware for Fisher Scientific accumet AB150, AB200, and AB250 series meters which may become available.
  • Page 28 3. Select “Install from a list or specific location (Advanced)” as shown in figure above and then click “Next”. 4. Select “Search for the best driver in these locations” and enter the file path in the combo-box (“C:\Accumet_AB_USB_Driver” as in figure below) or browse to it by clicking the browse button.
  • Page 29 5. If Windows XP is configured to warn when unsigned (non-WQHL certified) drivers are about to be installed, the message dialogue shown in figure below will be displayed unless installing a Microsoft WHQL certified driver. Click on “Continue Anyway” to continue with the installation.
  • Page 30 6. The screen shown in the figure below will be displayed as Windows XP copies the required driver files. 7. Once the installation is successful as shown in the figure below. Click “Finish” to complete.
  • Page 31 USB Driver Installation Guide For Windows 7: To install the USB device driver under Windows 7, follow the instructions below. This will allow your computer to recognize the instrument. 1. Obtain the USB driver file (via web download, disc, email, etc.) and save on your computer in a location that can be easily found.
  • Page 32 4. Right click on the other devices (USB VCOM Port) to bring up a menu as shown below. From the Displayed menu, select “Update Driver Software…”. 5. Select “Browse my computer for driver software” and enter the file path in the combo-box or browse to it by clicking the browse button.
  • Page 33 6. Once the file path has been entered in the box, check “Include subfolders” box and click “Next” to proceed. 7. If necessary, click on “Install this driver software anyway” to continue with the installation if you receive Windows 7 security pop up window.
  • Page 34  The following window appears when installation is in progress.  When the installation has finished completion screen is displayed as shown below.
  • Page 35 8. Click “Close” and go back to the Device Manager Window.  The below figure shows the successful installation of USB Driver for Accumet AB meter. Note: If the USB cable is removed and connected to a new (different) USB port, please repeat the driver installation procedure for the new USB port.
  • Page 36 Boot Loading Procedure: The software version of your instrument is displayed on the screen during power on. Check www.fishersci.com/accumet or email accumet@fishersci.com to see if new software is available. As new releases become available, the boot loading procedure described here is the process by which you can upgrade your Fisher Scientific accumet®AB150, AB200, and AB250 series meter to have the latest firmware.
  • Page 37 5. The instrument will beep twice and a pop-up window will appear showing contents of a new drive named “ACCUMET AB” on your computer similar to the one seen below. This folder represents the hard drive of the instrument. The left soft key can be released as soon as this window is visible.
  • Page 38 6. Similar to the method for replacing a file that resided on a thumb drive, you will need to delete the firmware.bin file as seen in the pop-up window and replace it with the new file. Highlight “firmware.bin” and delete this file. 7.
  • Page 39: Specifications

    8. Once the file is successfully transferred, close the window. 9. Remove the power cord and restart the meter as you would normally. The new software version will be briefly visible during power on. Specifications We reserve the right to make changes, improvements and modifications to the specifications listed here.
  • Page 40 AB250 Only Range 0.001 to 19999 (±2000 mV) Resolution 0.001 / 0.01 / 0.1 / 1 (automatic) Units ppm, mg/L, molar 0.5 % full scale (monovalent ion) Accuracy 1 % full scale (divalent ion) 2 to 6 points from one of following groups; ...
  • Page 41 Salinity AB200 Only Range 0 to 80.0 ppt Resolution 0.01 / 0.1 ppm; 0.001 / 0.01 / 0.1 ppt Accuracy ±1 % full scale Resistivity AB200 Only Range 2.000 Ω to 20.0 MΩ Resolution 0.01 / 0.1Ω; 0.001 / 0.1 kΩ; 0.01 MΩ Accuracy ±1 % full scale Other...
  • Page 42: Replacements And Accessories

    Weight Benchtop: 720 g (without electrode plate) Size Benchtop: 7 cm (H) x 18.5 cm (W) x 17.6 cm (D) Regulatory and Safety CE, TUV 3-1, FCC Class A Power Rating DC Input: 9 VDC 2A Vibration: shipping/handling per ISTA #1A Shock and Vibration Shock: drop test in packaging per ISTA #1A Enclosure...
  • Page 43 AB200 pH/CON meter only 13-636-AB200A AB200 pH/CON KIT, includes 13-620-631 pH electrode and 13-636-AB200 13-620-100 conductivity cell ® Fisher Scientific accumet benchtop stirring probe 13-620-BSP Replacement paddle for 13-620-BSP 13-620-RP pH/ATC electrode, double junction, plastic body, refillable 13-620-631 pH electrode, double junction, glass body, refillable...
  • Page 44: Warranty

    Warranty The Fisher Scientific Company (“Fisher”) warrants to the direct purchaser that the accumet meters and accumet, accuTupH, and accupHast, electrodes will be free from defects in material or workmanship for a specified warranty period. During that period, Fisher will repair or replace the product or provide credit, at its sole option, upon prompt notification and compliance with its instructions.
  • Page 45: Notice Of Compliance

    Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the symbol on the left. Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State and this product should be disposed of or recycled through them.
  • Page 46: Declaration Of Conformity

    Singapore Hereby declares that the following products rated 100-240 VAC, 50/60 Hz, 2A:  Fisher Scientific accumet AB150  Fisher Scientific accumet AB200  Fisher Scientific accumet AB250 Equipment Class: Measurement, control and laboratory, EMC Class A Conforms to the following directives and standards:...
  • Page 48 Fisher Scientific accumet Basic (AB) ® Benchtop Meters ● AB150 ● AB200 ● AB250 For technical assistence contact your Fisher Scientific representative or visit: www.fishersci.com/accumet or email accumet@fishersci.com Thermo Fisher Scientific (Australia) +1300-735-292 thermofisher.com.au Fisher Scientific GmbH (Austria) 0043-800-208840 at.fishersci.com...