P2.0 View Electrode Diagnosis Program

This program allows the electrode parameters to be checked for diagnostic purposes.
Depending on the last display mode, the displayed electrode information will vary. If pH was
the display mode, the electrode offset value in mV and slope in percent will be shown. If mV
was the display mode, the electrode offset value in mV will be shown.
1. With the meter powered off, press and hold the MODE
key and press and release the ON/OFF key. The display
will show the "SETUP" indicator.
2. Press the  or  arrow key to scroll through the
programs until "ELE" is shown in the upper display and
"P2.0" is shown in the lower display. The electrode icon
will also appear.
3. Press the ENTER key to access the program P2.0. The
lower display will show "P2.1".
4. The information shown in program P2.0 will depend on
the mode the meter was in prior to powering the meter
off:
a. In the pH measurement mode:
i. P2.1 shows the mV offset of the electrode.
Press the ENTER key to go to P2.2.
ii. P2.2 shows the slope in % of the electrode.
b. In the mV measurement mode:
i. P2.1 shows the mV offset of the electrode.
5. To exit P2.0 program press the ENTER key until "ELE"
is shown in the upper display and "P2.0" is shown in the
lower display.
6. To go to the next program, press the  or  arrow key
or to return to the measurement mode press the
CAL/MEAS key.