15.
PARAMETER SETUP: DATA STORAGE CRITERIA
The Data Storage Criteria setting is available from any parameter Setup. Use this option to select which
criteria to include when saving data into memory.
To Set Data Storage Criteria:
1. Select Touch here to edit next to Data Storage Criteria.
2. Select ON or OFF for each detail.
3. For Data Logging Interval select MANUAL, STABLE, or TIMED.
MANUAL: Saves data to memory only when Log Data is pressed.
STABLE: Saves data to memory only when the STABLE indicator appears.
TIMED: Saves data to memory automatically at user determined intervals (3 sec minimum).
4. Select OK to accept your selections.
21