4.1.1 Supplies needed
When protection box without optional power supply is used:
M12 extension cable, supplied either as a:
•
a) standard, molded connector M12 extension cable, or
b) kit for field-assembled M12 extension cable, including:
- 30 m [100 ft.] extension cable (VSM P/N: 6001-11502 [Belden
- M12 connector
When protection box with optional power supply is used:
30 m [100 ft.] extension cable (VSM P/N: 6001-11502 [Belden
4.1.2 Tools needed
fish-tape (to pull M12 extension cable inside the protective conduit)
•
wire-stripper
•
assorted miniature flat screwdriver
•
4.2 Step by step installation
4.2.1 Connection of the SMA triaxial cable carrying signal input
Figure 24 : Triaxial cable connected to a LIN-300 Series module
32
LIN-300 Series Modules_Installation - Rev.4
®
#9940])
®
#9940])
1. Insert the SMA male connector and engage
by turning clockwise.
Recommended SMA connection
torque: 1.7 Nm [15 in-lb]
2. Insert the tooth lock washer, the round lug,
the spring lock washer and the nut onto the
grounding bolt. Screw the nut firmly to ensure
the grounding connection.
Recommended grounding connec-
•
tion torque: 2.8 Nm [25 in-lb]
The ground lug must not come in
•
contact with neither SMA con-
nector nor M12 connector.