Whenever it is likely that safety protection has been impaired, make the measuring chain inoperative and • secure it against any unintended operation. Have the LIN-300 Series modules serviced only by qualified service personnel. • Safety and electrical symbols that appear in this manual and on the material: •...
Due to the calibration process, the length of the integral cable may vary lightly from one sensor to another. • Minimum length is 5.5 m [18 ft.]. Protection boxes for the LIN-300 Series signal conditioners must be installed within this range.
2. Adjust the sensor at mid-range distance from the target, using a gauge between sensor face and target. Refer to the specifications of TARGET the LIN-300 Series module used 3. Set the sensor in place by tightening jam nuts at recommended torque. When adjustment...
2. Place a temporary cable between the sensor casing and insulated bearing to create a short-circuit path. 3. Complete the measuring chain by connecting the sensor to its LIN-300 Series conditioner and apply power to the conditioner. Take a measurement of the 4-20 mA output signal.
3.1 Installation of the 10X6X3 ABS Protection Box 3.1.1 Preliminary considerations The ABS protection box is a plastic enclosure which can house a single LIN-300 Series conditioner. Outer dimensions are 254 x 152 x 76 mm (10 x 6 x 3 in.).
The extension cable can be assembled on site, using 4-conductor shielded cable and an M12 connector, • or a standard molded M12 cable assembled can be used. The standard cable offered by VibroSystM is a molded M12 cable, 30m (100 ft.) long, and is rated for a maximum operating temperature of 80 °...
SMA connector nor M12 connector. Figure 27 : Power and Signal output cable (right) connected to a LIN-300 Series module On some models of molded cables, the M12 connector is equipped with an internal ratchet element preventing unintentional loosening of the compression nut. M12 connectors with this feature produce a clicking sound when turning the compression nut.
(50 Hz - 60 Hz) DC Input voltage:90V - 350V Figure 28 : LIN-300 Series modules inside a protection box with optional power supply Caution The protective conductor terminal must be connected to a grounding point in compliance with local regulations. Grounding is essential for safety purposes as well as to provide better efficiency against ESD and EMI perturbations.
3. a green or red pulsation. b) the output signal value of the LIN-300 Series module should be in the range of 3.8 to 20.6 mA. The output is a linear current throughout the measuring range. Refer to the LIN-300 technical specifications for values needed for precise slope-intercept form calculations.
4.4 LIN-300 Series Modules - General specifications Operation Matching sensor PCS-202SB • ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- LIN-302S-6I-0,3/2,3 Measuring range 0.3 to 2.3 mm [12 to 91 mils] • Output 4 to 20 mA • Sensitivity 8 mA/mm [0.202 mA /mil] • Accuracy < 3% of reading at mid-range •...