10-2. GLP report

To print the GLP report to the AD-8121B, select the function setting "info 1" and
"pUse 1" for the balance, and use MODE 3 of the printer.
To transmit the GLP report to a personal computer, select the function setting "info
2" and "pUse 0".
The serial number attached to the lateral side of the balance may have a form
"#Q1234567" (# = I, E and so on). In this case, the first letter "#" is ignored as a
GLP data and output data has a form "Q1234567".
Calibration report
This function is not available after sealing the calibration (CAL) switch.
1. Perform calibration according to "7-1. Calibration
using a weight".
end
2.
appears when the calibration has
been completed.
3. glp
is displayed and calibration report is
output.
4.
end
appears again. Remove the weight
and press the
weighing mode.
AD-8121 format "info 1"
'
MODEL
EK-1200i
S/N
Q1234567
ID
DATE
2011/11/14
TIME
02:53:21
CALIBRATED(EXT.)
CAL.WEIGHT
+1000.0
SIGNATURE
- - - - - - - -
MODE
key to return to the
Manufacturer
A & D
Serial number
ABCDEF
g
Calibration weight
Column for
:
Space, ASCII 20h
CR: Carriage return, ASCII 0Dh
LF:
Line feed, ASCII 0Dh
Operation of Calibration
(See "7.CALIBRATION")
Returns to the weighing mode
General format "info 2"
Model
ID number
Date
Time
Calibration
executed
signature
26
end
glp
To output
calibration report
end
`~~~~~~~~~~A~&~D<CRLF>
MODEL~~~EK-1200i<CRLF>
S/N~~~~~Q1234567<CRLF>
ID~~~~~~~~ABCDEF<CRLF>
DATE<CRLF>
<CRLF>
TIME<CRLF>
<CRLF>
CALIBRATED(EXT.)<CRLF>
CAL.WEIGHT<CRLF>
~~~~~~+1000.0~~
<CRLF>
g
SIGNATURE<CRLF>
<CRLF>
<CRLF>
---------------~<CRLF>
<CRLF>
<CRLF>