shows an error that pops up if the computer cannot communicate with the EVM. If you receive
this error, first ensure that the USB cable is properly connected on both ends. This error can also occur if
you connect the USB cable before the USB DIG Platform power source. Another possible cause for this
error is a problem with the USB Human Interface Device Driver on the computer you are using. Make sure
that the device is recognized when the USB cable is plugged in; this action is indicated by a Windows-
generated confirmation sound.
6.2
Using the BUF12800 Software
2
6.2.1
I
C Address Selection
As mentioned previously (refer to
BUF12800.
Figure 15
between the BUF12800EVM and the software. Without jumper JMP6 and the software address button
configured correctly, the software cannot communicate with the BUF12800 device.