Section 3
Configuration
RS485 Address
Back Up System Type
Cold Excursion
Warm Excursion
Reset Excursion
3-4
TSC Series
If the freezer is configured for RS-485 communications, it will need to
have a unique identification address. This address is set through the
Configuration mode.
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until RS485ADDR is displayed in the message
center.
3. Press Enter. The message center will display 485 ADDR XX.
4. Press up/down arrow to select the appropriate address for the freezer
(1 - 24).
5. Press Enter to save.
This function, which is only available on freezers with the optional back
up system (BUS), allows the user to select which type of gas is injected into
the freezer chamber. The options are CO
1. Press the Mode key until the Configuration indicator lights.
2. Press the right arrow until BUS TYPE CO2 or BUS TYPE LN2 is
displayed in the message center.
3. Press up/down arrow to toggle between the two display selections.
4. Press Enter to save.
This function displays the coldest temperature recorded by the control
probe.
This function displays the warmest temperature recorded by the control
probe.
This function resets the cold and warm excursions.
and LN
.
2
2
Thermo Scientific