Replace Battery(S) - Thermo Scientific TSC Series User Manual

-40c and -86c
Table of Contents

Advertisement

Replace Battery(s)

All Models Except 3 cu ft
3 cu ft Models
Thermo Scientific
The following instructions describe the battery replacement procedure for
specific models.
1. Open the lower panel door by grasping the handle and pulling.
2. Locate the battery power switch (Figure 5-1). Turn the battery power
switch to the Off position (O).
3. Remove the four screws holding the recorder bezel to gain access to the
battery.
4. Remove the three nuts securing the battery bracket. See Figure 5-1.
5. Remove the bracket and old battery. Discard properly. Install the new
battery and secure.
6. Reconnect the battery (red to positive and black to negative).
7. Replace the recorder bezel.
8. Turn the battery power switch to Standby mode ( ).
9. Close lower panel door.
1. Open the lower panel door by locating the handle on the underside of
this door and pulling.
2. Locate the battery power switch (Figure 5-2). Turn the battery power
switch to the Off position (O).
3. Remove the two nuts securing the battery bracket. See Figure 5-2.
5. Remove the bracket and old battery. Discard properly. Install the new
battery and secure.
6. Reconnect the battery (red to positive and black to negative).
7. Replace the recorder bezel.
8. Turn the battery power switch to Standby mode ( ).
9. Close lower panel door.
Note For a consistent and dependable charge, replace the battery every 2
years. Replacement batteries must be rechargeable and are available from
Thermo. Refer to the parts list for stock number and description of the
replacement batteries (P/N 400159). Dispose of the used batteries in a safe
manner and in accordance with good environmental practices.
Section 5
Maintenance
s
TSC Series
5-3