Certifications
Declaration of Conformity is available upon request.
Safety Specifications
Indoor Use Only
Altitude - up to 2,000 meters
Temperature - 5°C to 43°C
Humidity - Maximum RH 80% for temperatures up to 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
Installation Category II
1
2
Pollution Degree 2
Class of Equipment I
Intended Use
This product is intended for use as a General Purpose Laboratory Freezer for storing samples or inventory between -50 and -86°C for units rated at
-86°C and -10 and -40°C for units rated at -40°C .
This unit is not intended for use in an explosive environment, nor to be used for the storage of flammable inventory. This unit is not intended for
use in a Class II medical application as defined by Title 21 of the Federal Code of Regulations.
Fluorinated Greenhouse Gases
Compliant with REGULATION (EU) No 517/2014 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL on fluorinated greenhouse gases.
This product contains fluorinated greenhouse gases in a hermetically sealed system. This product contains foam blown with fluorinated
greenhouse gas, R-245fa. If a leak in the sealed system is detected, the operator shall repair without undue delay.
For ULT models rated 230V/50Hz, intended for import into the EU, the following sizes are designed for use with the following amounts of fluorinat-
ed greenhouse gases:
For -86°C rated units
Size (ft³)
Refrigerant
3
R-404a
12.7
R-404a
17
R-404a
20
R-404a
For -40°C rated units
Size (ft³)
Refrigerant
3
R-404a
12.7
R-404a
17
R-404a
20
R-404a
______________________________________________________________________
1 Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to withstand safely. It depends
on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II which is the category used for instruments in
installations supplied from a supply comparable to public mains such as hospital and research laboratories and most industrial laboratories, the expected
transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
2
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2 assumes that normally only
non-conductive pollution such as dust occurs with the exception of occasional conductivity caused by condensation.
Thermo Scientific
1st Stage
Amount (kg)
GWP
0.227
3 922
0.680
3 922
0.680
3 922
0.680
3 922
1st Stage
Amount (kg)
GWP
0.425
3 922
0.539
3 922
0.595
3 922
0.652
3 922
2nd Stage
Refrigerant
Amount (kg)
R-508b
0.173
R-508b
0.425
R-508b
0.439
R-508b
0.454
Section 7
Specifications
GWP
13 396
13 396
13 396
13 396
TSC Series
7-5

Advertisement