Alarms - Thermo Scientific TSC Series User Manual

-40c and -86c
Table of Contents

Advertisement

Thermo Scientific
Section 4
The TSC Series freezer alarm system is shown below. When an alarm is
active, a message appears in the LED message center. Press the Silence key
to mute the audible alarm for the ringback period. The visual alarm
continues until the freezer returns to a normal condition. The alarms are
momentary alarms only. If an alarm condition occurs and then returns to
normal, the freezer automatically clears the alarm condition and the
message center.
Table 4-1. Alarms
Description
No alarm condition exists
Power Failure
High Temperature Alarm
Low Temperature Alarm
Door Ajar
Low Battery*
Low BUS Battery (optional)
Control Probe Failure
Heat Exchanger Probe Failure **
Condenser Probe
Sample Probe Failure (optional)
High Stage System Failure **
Condenser Hot Condition
Wrong Power
Micro Board Failure
All alarm delays and ringback times are +30 seconds.
*The automatic battery test runs immediately on power-up, then every 8 hours thereafter.
** -86C units only

Alarms

Message
SYSTEM OK
POWER FAIL
TEMP IS HIGH
TEMP IS LOW
DOOR IS OPEN
LOW BATTERY
LOW BUS BATT
CNT PRB FLT
HSHX PRB FLT
COND PRB FLT
SMPL PRB FLT
HS SYST FAIL
HOT CONDENSR
WRONG POWER
MICRO FAIL
Delay
Ringback
----
----
1 min.
15 min.
1 min.
15 min.
1 min.
15 min.
1 min.
15 min.
1 min.
8 hours
1 min
15 min.
1 min.
15 min.
1 min.
15 min.
1 min.
15 min.
1 min.
15 min.
1 min.
15 min.
1 min.
none
0 min.
none
0 min.
15 min.
Relay
----
Yes
Yes
Yes
No
No
No
Yes
No
No
No
YES
No
YES
YES
TSC Series
4-1

Advertisement

Table of Contents