10.
Verify that all buttons on the unit work properly. Refer to the
Quick Guide for their applications, see Figure 10-4.
Short Press
Power On
Changes Displays /
Selects Parameters
Event Mark
Changes
Values / Contrast
Sound ON / OFF
Figure 10-4: Quick Guide
11.
Verify the CO
Calibration Check process functions. Follow the
2
procedure as described below. When completed, return to the
Measuring Mode screen.
CO
Calibration Check
2
Caution: Do not check CO
BTPS setting is ON (factory default). This mode corrects the CO
value for BTPS. The calibration check mode disables this
correction.
In order to ensure accuracy, a CO
a year. Calibration gas and a FilterLine are needed for this procedure.
Start the process from the Interface Test Screen (refer to Table 4-2) as
follows in Table 10-1.
Note: Connect the FilterLine to the monitor before starting CO
Table 10-1: CO
Objective
Access CO
short press (x3)
2
Calib.
Change option
short press
to start.
Section 10: Performance Verification
Long Press
Accesses
Menus
Home /
Erase Trend
Quick Scroll
Accesses
Alarm Menu
Data Transfer
ON / OFF
values from the measuring mode if the
2
calibration check needs to be done once
2
Calibration Check
2
Action
Result
2
Calibration Check.
2
10-3