SECTION 3: CLEANING

3.1

CLEANING

3.1.1

Procedure

3.1
Cleaning
Always clean the outer surface of the monitor before servicing.
Warning: Always wear latex or surgical gloves when cleaning
or servicing the Handheld Capnograph monitor after hospital
use.
Warning: Do not allow liquids to drip inside the housing.
Warning: Do not clean internal components.
Warning: Do not use cleaning materials based on: phenols,
halogen producing compounds, strong organic acids, or oxygen
producing compounds.
1.
Unplug the unit from AC mains supply.
2.
With a damp cloth, gently wipe the outer case of the monitor with
a disinfectant basd on aldheyde, alcohol or quaternary ammonium
producing compounds.