Section 7: Disassembly Guide
7.8

UPDATING SOFTWARE VERSION

7-12
Caution: Observe ESD (electrostatic discharge) precautions when
disassembling and reassembling the monitor and when handling
any of the components of the monitor.
1.
Open the monitor as described in Section 7.2.
2.
Remove the EPROM from its socket (on the CO
extractor, refer to Figure 7-12.
3.
Insert the new EPROM.
Note: When replacing the EPROM, be sure it is positioned correctly.
Note: When operating the monitor for the first time after updating the
software version, the warm-up and the self-test periods will be
longer than usual.
Figure 7-12: Replacing EPROM
board) using an IC