SECTION 2: REQUIRED EQUIPMENT

2.1

REQUIRED EQUIPMENT

2.1
Required Equipment
You will need the following equipment to disassemble, replace parts,
check, adjust, or calibrate the monitor.
Note: For ordering spare parts, contact your local distributor.
For Disassembly and Reasssembly:
• Phillips-head screwdriver (medium)
• IC Extractor for PLCC socket
• Threadlocker glue e.g. Loctite® 222 super screw lock
For Calibration, Checks and Adjustments:
• Calibration Gas (5% CO
• FilterLine or FilterLine H
• Leak Test Jig
• Vacuum manometer with a range of 0-500 mBar
e.g. MPB-3200-DM, Digital Manometer MicroPneumatic Logic, Inc.
• Communication Adapter Kit (only necessary to verify print and
PC functions)
• Seiko DPU-414 printer (only necessary to verify print functions)
• External power supply
• Handheld Capnograph Operator's Manual
• Flow Meter e.g. Top Track Mass Flow Meter,
model: 822-13-OV1-PV1-V1, Sierra Instruments, Inc.
Warning: Observe ESD (electrostatic discharge) precautions
when handling, adjusting or performing any procedure with the
monitor internal components.