4.3.1

Default Alarm Limits

The service technician can change the default alarm limits for either
patient mode (Adult or Neonatal).
The monitor has the following alarms with adjustable level settings:
• EtCO
high and low levels (mmHg)
2
• FiCO
high level (mmHg)
2
Refer to Table 4-3 for changing the settings of the above mentioned
parameters.
Warning: Changing default settings will permanently change the
user's start up value and should be verified with the responsible
clinical personnel.
Table 4-3: Changing Default Alarm Limits
Objective
To access the Alarm
Limits Menu after
entering last password
digit
To change the patient
mode
To access a new
parameter
To change the value
To reset and return to
measuring mode
To exit service mode
and save event
messages
Action
Response
short press
short press
short press
short press
long press
Turn off the
monitor.
Section 4: Service Mode
4-7