Mallinckrodt Nellcor NPB-70 Service Manual page 27

Handheld capnograph
Table of Contents

Advertisement

Table 5-1: Troubleshooting Guide (Continued)
Symptom
Probable Cause
The unit does not
The wall socket is not
turn on when the
receiving power or is
ON/OFF button
defective.
is switched ON
when the
monitor is
connected to
mains power.
The AC adapter is
defective.
A CO
has failed.
One or more
The keypad is defective.
buttons on the
upper or lower
keypad do not
work.
A CO
has failed.
No information
The contrast default
is displayed on
value is set too low or
the LCD.
high.
An LCD component has
failed.
A CO
has failed.
LCD Backlight
The Power
does not come on
Management option is
when the
set at LOW .
monitor is placed
in a dark
environment.
An LCD component has
failed.
A CO
has failed.
Corrective Action
Check the wall socket
for power supply or
mechanical defects. If
necessary, use a
different wall socket.
Replace with new AC
adapter.
Replace with a new
board component
2
CO
Inspect the keypad for
cracks, deformations
or other damage.
Replace with a new
keypad.
Replace with a new
board component
2
CO
Check the contrast
value in the Service
Mode using the
Contrast Cal. option
and correct
accordingly.
Replace with a new
LCD.
Replace with a new
board component
2
CO
Refer to the Handheld
Capnograph
Operator's manual to
change the Power
Management option
to NORMAL or
HIGH .
Replace with a new
LCD.
board component
Replace with a new
2
CO
Section 5: Troubleshooting
board.
2
board.
2
board.
2
board.
2
5-3