Keysight Technologies 34941A User Manual page 12

Rf multiplexer modules
Table of Contents

Advertisement

Electrical Isolation
You can configure each bank on the RF MUX modules to be either isolated or
chassis-grounded. The modules come with chassis-grounded metal shoulder
washers installed on all connectors in each bank of relays. If you want to isolate a
bank from the other banks and from chassis-ground, you must remove the five
metal washers in that bank and replace them with the provided plastic shoulder
washers.
The 34941A and 34942A are
shipped from the factory
chassis-grounded with metal
shoulder washers installed on all
connectors in each bank of relays.
Keysight 34941A-34942A User's Guide
To isolate a bank from other
banks and chassis-ground,
install the provided plastic
shoulder washers on the
connectors in the bank to be
isolated. When installing the
plastic shoulder washers, use
7 in-lbs of torque.
11