Specifications; Battery Replacement - Greenlee CMI-200 Instruction Manual

Clamp-on ammeter
Table of Contents

Advertisement

Available languages

Available languages

Specifications

Display: 3-3/4-digit LCD (3999 maximum reading) and a 40-segment bar graph
Auto Power Off: After 30 minutes of inactivity
Polarity: Automatic
Sampling Rate:
Numeric Display: 2 per second
Bar Graph Display: 20 per second
Over Range Indication: Leftmost digit will flash
Jaw Opening: 55 mm (2.17")
Overvoltage Category: Category III, 600 Volts
Operating Conditions: 4 C to 50 C (39.2 F to 122 F), 0% to 85% relative humidity
(noncondensing)
Elevation: 2000 m (6500') maximum
Indoor use
Storage Conditions: -20 C to 60 C (-4 F to 140 F), 0% to 75% relative humidity
(noncondensing)
Remove battery.
Pollution Degree: 2
Battery: 9-Volt battery (NEDA 1604, JIS 006P or IEC 6LF22)

Battery Replacement

Before opening the case, remove the jaw from the circuit and shut off the unit.
Failure to observe this warning can result in severe injury or death.
1.
Disconnect the unit from the circuit. Turn the unit OFF.
2.
Remove the screws from the back cover.
3.
Remove the back cover.
4.
Replace the battery (observe polarity).
5.
Replace the cover and screws.
16