Rack Mounting - R&S NGL-COM2A Getting Started

Hide thumbs Also See for NGL-COM2A:
Table of Contents

Advertisement

®
R&S
NGL200/NGM200
Figure 3-1: Operating positions
Positioning of instrument
The instrument must be positioned in a manner that allows you to discon-
nect the unit from the mains at any time and without restrictions.
3.4.2

Rack Mounting

The instrument can be installed in a 19" rack using the rack adapter R&S HZN96
(P/N 3638.7813.02). Proceed according to the installation instructions supplied
with the rack adapter.
Ambient temperature
Place the R&S NGL/NGM power supply in an area where the ambient tem-
perature is within +5 °C to +40 °C. The R&S NGL/NGM power supply is
fan-cooled and must be installed with sufficient space along the sides to
ensure free flow of air.
Getting Started 1178.8720.02 ─ 05
Putting into Operation
Setting Up the Instrument
15