ALARM PARAMETERS
5. On the Alarm Setting #2 screen, it is possible to
set each alarm. Press Apply key to save the input
value and setup. The display returns to the Tools #1
screen.
●Each setting
・Alarm Delay:
The function is that when the incubator is in the alarm state of Automatic set temperature or of Automatic
set CO
density, the alarm buzzer will sound after the alarm delay set time passed. Settable range: 0
2
minute~15 minutes, factory setting: 15 minutes.
Note: When the incubator is recovered from the alarm state within the alarm delay time, the buzzer
doesn't sound after the elapse of the alarm delay.
・Door Delay:
The function is that when the incubator is in the alarm state of door, the alarm buzzer will sound after the
alarm delay set time passed. Settable range: 1 minute~30 minutes, factory setting: 2 minutes.
Note: When the incubator is recovered from the alarm state within the door alarm delay time, the buzzer
doesn't sound after the elapse of the door alarm delay.
・Ring Back:
The function is that the alarm buzzer sounds again when the alarm state still continues after the alarm
delay set time passed even though the alarm buzzer was stopped by pressing Buzzer key. By holding and
sliding Ring Back slide key to the right, the Ring Back is turned to ON. Settable range: 1 minute~99
minutes, factory setting: 30 minutes.
Note: At Err01 (CO
gas cylinder empty), Err11・12 (CO
sensor error), Err18 (UV lamp failure) and Door
2
2
alarm, the alarm is not re-activated because the alarm itself is deactivated by pressing Buzzer key (refer to
page 78 and 79).
・Remote Alarm:
The function is that the remote alarm is continued even though the buzzer is stopped by pressing Buzzer
key. By holding and sliding Remote Alarm slide key to the right, the Ring Back is turned to ON (not in
conjunction with Buzzer key). Factory setting: ON.
6. (From procedure 4 and 5) Press Top key to return to the Top screen.