Download Print this page

Panasonic MCO-170AICUVH Series Operating Instructions Manual page 12

Co2

Advertisement

1. Outer door: The outer door is held to the frame with the magnetic seal. The door heater is installed in
the door panel. The door opening is reversible. Contact our sales representative or agent to change the
door hinge from left to right or vice versa.
2. Inner door: The inner door is made of tempered glass. However do not subject the glass to excessive
impacts.
3. Tray catches: Insert tray to fit the concave portion on chamber.
4. Fan cover: The fan cover serves as the inlet for circulating air. It is removable.
5. Duct: The duct for the path for circulating air. It is removable.
6. Fan (inside the duct): The fan is made from polypropylene resin. It can be disinfected in an autoclave.
1
7. UV lamp*
: This UV lamp does not generate ozone. Never look directly at the UV light. For replacement,
contact our sales representative or agent.
8. Humidity control bar: Reduce automatically dew condensation occurred by the effect of outside
environment and the frequency of door opening/closing.
The humidity control bar has bactericidal effects by plated surface of it. However, it is recommended to
replace the humidity control bar every 5 to 6 years to maintain bactericidal effects. (The duration
of bactericidal effects differs depending on the use environment.)
9. Humidifying pan cover: This cover prevents the UV light entering the chamber. Always use it. Using
without it may have a bad influence on the chamber temperature distribution and humidity recovery.
10. Humidifying pan: Fill the humidifying pan with sterile distilled water. Install the humidifying pan in a
longitudinal direction as its shorter side is placed in the back.
11. Door switch: Detects the door opening/closing and stops the fan and electromagnetic valve for CO
2
1
when the door is open. The UV lamp*
is also deactivated by the door opening.
2
12. Key hole*
: This is the hole to unlock with unlock key while outer door is locked by electric lock.
13. Switch cover: Prevent the accident of gas tube disconnected by the unexpected touch or power off.
14. Power supply cord cover plate: This plate is to prevent the power cord being come off.
15. Connecting port A for CO
gas pipe: Refer to page 20 for gas cylinder connection. Ensure that the
2
2
gas pressure is set at 0.03 MPa(G) (0.3 kgf/cm
(G), 4.3 psi(G)).
Note: When the optional MCO-21GC gas auto changer is installed, both ports A and B are available. Refer
to page 70 for gas auto changer.
1
16. Glow starter*
: The glow is started for the UV lamp. (Type FG-7P or FG-7PL)
17. Power switch: This is the main switch for the incubator. It also functions as an overcurrent breaker.
18. Sample air outlet: The sample air outlet also functions as an internal gas outlet. Normally, cover this
outlet with the sample air outlet cap.
19. USB port: Insert USB memory to export operations and alarms log. Refer to page 40~51.
Note: It is impossible to use USB memory which is required password input.
20. Access port: Place the silicon caps on both outside and inside of the port when the port is not being
used.
21. Remote alarm terminals: This terminal informs the alarm to remote location by connecting to external
alarm unit. Refer to page 15.
11

Advertisement

loading

This manual is also suitable for:

Mco-170aicMco-170aicuvMco-170aicuvhMco-170aicuv-pe