Co2

Advertisement

3. Input each parameter. Press Apply key to save
the input value. The display returns to the Menu
screen.
●Each parameter setting
・Temperature: Set value of chamber temperature.
o
Settable range: 0.0
C~50.0
Density: Set value of chamber CO
・CO
2
・High Limit: The high limit temperature alarm is different from the Automatic set temperature alarm (page
37), and it is independent temperature alarm. In case the chamber temperature exceeds the temperature
of the high limit temperature alarm, this alarm is activated.
o
Settable range: 20.0
C~53.0
・Auto-lock: Refer to page 67~69.
Refer to page 78~80 for detail of each alarm.
Note:
・When operating the incubator for the first time or after not using it for an extended period of time, operate
it for at least about 4 hours until the chamber temperature and the CO
chamber temperature to the desired temperature and setting the CO
setting to the desired CO
2
・Set the temperature of the high limit temperature alarm after the chamber temperature is stable at the set
value.
・Set the high limit temperature alarm to at least 5
4. On the Menu screen, press Back key to return to the Top screen.
o
C, factory setting: 37.0
density. Settable range: 0.0 %~20.0 %, factory setting: 0.0 %。
2
o
C, factory setting: 53.0
density.
o
C higher than the chamber set temperature.
o
C。
o
C。
sensor are stable after setting the
2
density to 0 %. Then change the
2
33