System Fuse Specifications - Hach BioTector B3500dw User Manual

Online toc
Table of Contents

Advertisement

5.2.4 System Fuse Specifications

Table 5 below summarizes the location and specification of the fuses used in BioTector. The fuse values
tabulated below can be changed depending on system selected options. The locations of the fuses are also
displayed in figure 9 above.
DANGER
Table 5 System Fuse Specifications
Location
Board
Number
Power and
Input/Output
81204350-01
Board
Motherboard
81204340_02
BioTector
Compressor
N/A
Start Signal
Cable
KEY
A:
Amperes
DIN:
German Institute for Standardization (Deutsches Institut für Normung e.V.)
F1-6:
Fuse Number
F:
Fast Acting (Fast Blow)
H:
High Interrupt
ID:
Identification
L:
Low Interrupt
mA:
Milli-amperes
N/A:
Not Applicable
PCB:
Printed Circuit Board
T:
Time Lag (Time Delay)
V:
Volts
*1:
(Optional) used with TDK power supply, type SWS300-24 only.
*2:
(Optional) if higher powered device are connected, for example Multi-channel Modbus / Profibus,
External Valve, External Relay.
BioTector contains electrical components operating under high voltages.
Contact may result in electric shock and severe or fatal injury.
All electrical work should be carried out by qualified electrical personnel only.
When any fuse replacement is required in the system, please refer to table 5
below and system specific drawings.
Interru
Type
pt
Rating
Miniature
H-250V
5x20mm
Miniature
L-250V
5x20mm
Miniature
L-250V
5x20mm
Material
Fuse
Number
F1
*1
F1
Ceramic
F2
F3, F4
F3, F4
*2
F2
F3
Glass
F4
F5
F6
Glass
N/A
230 V
115 V
Systems
Systems
Current
Current
Rating
Rating
T 2.00A
T 3.15A
T 3.15A
T 5.00A
T 1.00A
T 1.00A
T 500mA
T 500mA
T 1.00A
T 1.00A
F 1.60A
F 1.60A
F 2.00A
F 2.00A
F 2.50A
F 2.50A
F 3.15A
F 3.15A
F 2.50A
F 2.50A
T 200mA
T 200mA
Page 56